IV.<br />

OPERATION<br />

<strong>SRJ</strong>-7<br />

<strong>HRJ</strong>-7<br />

A. TO MEASURE RAD1 AT1 ON I NTENSI TI ES:<br />

1. TURN THE SELECTOR SWI TCH TO THE ON POSITION AND WAI T FIVE SECONDS FOR<br />

THE ELECTROMETER TUBE FILAMENT TO WARM UP.<br />

2. TURN THE SELECTOR SWITCH TO THE SET POSITION AND ADJUST THE ZERO<br />

CONTROL UNTIL THE METER READS EXACTLY ZERO.<br />

3. TURN THE SELECTOR SWITCH TO THE 'IXI" POSITION. IF THE METER READS<br />

OFF SCALE, THE RADIATION PRESENT IS GREATER THAN 50 MR/HR IN THE<br />

CASE OF THE <strong>SRJ</strong>-7 (STANDARO RANGE), AND THE SELECTOR SWITCH SHOULD<br />

BE TURNED TO EITHER THE I1X1o1' OR<br />

l'X1OO1r POSITION.<br />

4. THE INSTRUMENT IS NOW READY FOR READING GAMMA RADIATION INTENSITY.<br />

TO READ GAMMA AND BETA SI MULTANEOUSLY, SLIDE BACK THE ALUMINUM<br />

SCREEN BY PULLING UP RIGHT HAND TAB "GI1 IN THE HANDLE.<br />

BETA AND GAMMA SIMULTANEOUSLY,<br />

ACETATE<br />

TO READ ALPHA,<br />

SLIDE BACK BOTH THE ALUMINUM AND<br />

FI LTERS BY PULLING UP BOTH TABS IN THE HANOLE.<br />

CAUT I ON<br />

WITH BOTH FILTERS BACK, THE ALPHA SCREEN IS EXPOSED AND<br />

CAN BE VERY EASILY DAMAGED. A VERY DELICATE TOUCH MAY<br />

CAUSE IT TO RIP.<br />

5; THESE INSTRUMENTS ARE CALf BRATED IN INTERNATIONAL ROENTGENS CORRECT-<br />

ED TO STANDARD TEMPERATURE AND PRESSURE (32OF. AND 29.92 INCHES HG.)<br />

WHEN USED AT 71 .6OF. (22°c)oAND 29.92 INCHES HG. (760 MM. HG.) FOR<br />

TEMPERATURE OTHER THAN 71.6 F. AND PRESSURES OTHER THAN 29.921' HG.,<br />

CORRECTION FACTOR<br />

FOR<br />

MUST BE APPLIED TO THE METER READING TO COMPENSATE<br />

IONIZATION CHAMBER AIR DENSITY CHANGES.<br />

CORRECTION FACTOR = 459.7 T°F. 29.92<br />

531 .3 P. INCHES HG.<br />

EXAMPLE: A<br />

IS 25 MR/HR ON THE XIO RANGE = 250 MR/HR. THE AIR<br />

TEMPERATURE IS 90 F. AND THE BAROMETRIC PRESSURE IS<br />

29.131l HG.<br />

459.7 t goo<br />

--- 549.7<br />

531 .3 531 .3<br />

29*92 1 .027<br />

29.13 -<br />

RADIATION MEASUREMENT IS TAKEN AND THE METER READING<br />

=I .035<br />

CORRECTION FACTOR =1 .035 x 1.027 = 1 .063<br />

CORRECTED READING = 250 MR/HR x i.063= 266 MR/HR.<br />

A<br />

THE EFFECTS OF AIR TEMPERATURE AND PRESSURE MUST BE TAKEN<br />

INTO<br />

- 4 -