4M- 10/58<br />

Bullefin No. 744<br />

INSTRUMENTATION FOR NUCLEAR RESEARCH<br />

IONIZATION CHAMBERS (MODELS IC-1 <strong>and</strong> IC-2)<br />

CONNECTOR<br />

I<br />

APPLICATION: These ionization chambers are used for detection <strong>and</strong><br />

measurement of Beta <strong>and</strong> Gamma radiation or Gamma radiation only. They<br />

can be installed in “hot” locations in a laboratory or reactor installation<br />

where it is desirable to monitor for control, health, or safety purposes.<br />

Model IC-1 has approximately one-half of the wall area cut away in the<br />

form of four windows which are fitted with .005” thick cellulose acetate.<br />

It is sensitive to soft radiation <strong>and</strong> is used for the detection <strong>and</strong> measurement<br />

of Beta <strong>and</strong> Gamma radiation. Model IC-2 is of solid wall construction<br />

<strong>and</strong> is used for the detection <strong>and</strong> measurement of Gamma radiation.<br />

CONSTRUCTION: These chambers consist of a bakelite tube 5” in diameter<br />

x 18%’’ long, with bakelite end plates. One end plate holds the high voltage<br />

input <strong>and</strong> signal output connectois. At the other end of the chamber, a<br />

space is provided to hold a dessicant to assui’e reliable operation under<br />

conditions of high humidity. Silica jell is used as the dessicant <strong>and</strong> is<br />

supplied in a cloth bag suitable for inseltion in the cavity provided.<br />

The collectoi, is an aluminum tube Yi’’ in diameter x 17%” long, <strong>and</strong><br />

is supported by Teflon insulation .at the back end, <strong>and</strong> by polystyrene<br />

insulation at the connector end. Guard rings are pi’ovided at either end to<br />

fuither reduce leakage. Leakage resistance is 10:’ ohms or higher to eithei.<br />

ground or high voltage.<br />

The inside of the 5” cylinder <strong>and</strong> the inside of the end plates are coated<br />

with aquadag to form a conducting surface. The aquadag coating is connected<br />

to the high voltage input connector. The electrical continuity resistance<br />

of the aquadag coating is held to less than 5000 ohms.<br />

Cai.eful craftsmanship thi.oughout assembly <strong>and</strong> testing insures<br />

that i.igid insulation specifications are maintained<br />

INQUIRIES INVITED<br />

In addition to the types<br />

SPEClFlCATlONS :<br />

OPERATING VOLTAGE RANGE: 275 to 325 volts D.C.<br />

described. Technical Associates<br />

- SENSITIVITY: Less than 1 mr/hr with micro-micro ammeter<br />

is prepared to supply ionization<br />

on 0.1 volt 10-11 ampere range.<br />

chambers for special purposes to<br />

* ENERGY DEPENDENCE: Small corrective factor below 0.3 mev;<br />

customer specifications.<br />

flat above 0.3 mev.<br />

* OIMENSIDNS: 5” dia. x 20” long. WEIGHT: Net 3 Ibs. Shipping 6 Ibs.<br />

TA<br />

TECHNICAL ASSOCIATES<br />

140 WEST PROVIDENCIA AVENUE BURBANK, CALIFORNIA