03.05.2014 Views

Instructions - OFI Testing Equipment, Inc.

Instructions - OFI Testing Equipment, Inc.

Instructions - OFI Testing Equipment, Inc.

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>OFI</strong> <strong>Testing</strong> <strong>Equipment</strong> – 146-00 <strong>Instructions</strong> – Workover and Completion Fluids Test Kit Page 1 of 17<br />

http://www.ofite.com<br />

INSTRUCTIONS<br />

WORKOVER AND COMPLETION FLUIDS TEST KIT<br />

PART No. 146-00<br />

Individual Test Procedures for the <strong>OFI</strong>TE Workover Kit are listed below:<br />

<strong>Instructions</strong><br />

ALKALINITY and LIME CONTENT<br />

WATER BASED DRILLING FLUIDS<br />

Alkalinity is the acid neutralizing power of a substance. Alkalinity measurements in drilling fluid testing may be made<br />

on the whole mud (designated with a subscript m) or on the filtrate (designated with a subscript f). The data collected<br />

can also be used to estimate the concentrations of hydroxyl (OH – ), carbonate (CO 3 -2 ) and bicarbonate (HCO 3 ) ions in<br />

the drilling fluid. Knowledge of the mud and filtrate alkalinity is important in many drilling operations. Mud additives,<br />

particularly some organic deflocculants, require an alkaline environment in order to function properly. Alkalinity<br />

arising from hydroxyl ions is generally accepted as being beneficial while alkalinities resulting from carbonates or<br />

bicarbonates may be detrimental to mud performance.<br />

The ions that are primarily responsible for filtrate alkalinity are the hydroxyl (OH – ), carbonate (CO 3 -2 ) and bicarbonate<br />

(HCO 3<br />

– ) ions. The carbonates can change from one form to another by changing the pH of the solution. Other<br />

inorganic ions such as borate's, silicates, sulfides and phosphates may also contribute to the alkalinity. It is important to<br />

realize the following calculations are only estimates of the concentrations of the reported ionic species based on<br />

theoretical chemical equilibrium reactions. The composition of mud filtrates is often so complex that the interpretation<br />

of alkalinities may be misleading. Any particular alkalinity value represents all of the ions which will react with the acid<br />

within the pH range over which that particular value was tested. Anionic organic thinners and filtrate reducers<br />

contribute to a large portion of the M f alkalinity value and may also mask the endpoint color change and render the test<br />

highly inaccurate in muds treated with organic thinners. For simple bentonite-based mud systems containing no organic<br />

thinners, the Phenolphthalein (P f ) and the methyl orange (M f ) alkalinities may be used as guidelines to determine the<br />

presence of carbonate/bicarbonate contamination and the treatment necessary to alleviate the problem. If organic<br />

thinners are present in large amounts, the conventional P f /M f test is suspect, and the P 1 /P 2 method should be used<br />

instead.<br />

<strong>OFI</strong>TE • 11302 Steeplecrest Dr. • Houston, TX 77065 USA • Phone (832) 320-7300 • http://www.ofite.com<br />

Version 1.1<br />

Date 03-04-03


<strong>OFI</strong> <strong>Testing</strong> <strong>Equipment</strong> – 146-00 <strong>Instructions</strong> – Workover and Completion Fluids Test Kit Page 2 of 17<br />

<strong>Equipment</strong> - Water Based Drilling Fluids:<br />

#147-16 Pocket pH Sensor (Optional)<br />

#153-26 Titration Dish, Polyethylene<br />

#153-28 Stirring Rod, Polyethylene<br />

#153-29 Syringe, glass tip, 2 ml<br />

#153-34 Pipette, Serological, 1 ml x 1/100 ml, glass<br />

#153-40 Pipette, Serological, 10 ml x 1/10 ml, glass<br />

Reagents:<br />

#220-01 Phenolphthalein Indicator sol'n, 8 oz<br />

#230-08 Sulfuric Acid sol'n, N/50, 8 oz<br />

#240-02 Methyl Orange Indicator sol'n, 8 oz<br />

Procedure - Filtrate Alkalinity, P f, M f<br />

1. Measure one or more milliliters of filtrate into the titration dish.<br />

2. Add two or more drops of Phenolphthalein Indicator solution. If the solution turns pink, add N/50<br />

Sulfuric Acid, drop by drop from the pipette, while stirring, until the pink color just disappears. If the<br />

filtrate is so colored that the end point cannot be seen, use a pH meter and titrate until the pH of the<br />

solution drops to pH 8.3, and this will be the end point.<br />

3. Report the Phenolphthalein Alkalinity of the filtrate, P f , as the number of milliliters of N/50 Sulfuric Acid<br />

required per milliliters of filtrate.<br />

4. To the same sample which was titrated to the P f endpoint, add two or three drops of Methyl Orange Indicator<br />

solution. Add the N/50 Sulfuric Acid drop by drop from the pipette while stirring until the color of the indicator<br />

changes from yellow to pink. The end point may also be taken when the pH of the sample drops to pH 4.3 as<br />

measured with a pH meter (more accurate).<br />

5. Report the Methyl Orange Alkalinity of the filtrate, M f , as the total milliliters of N/50 Sulfuric Acid per<br />

milliliters of filtrate required to reach the Methyl-Orange end-point. This also includes the amount of acid used<br />

to reach the Phenolphthalein, P f , end-point.<br />

Calculation, P f , M f (Estimation of Hydroxide (OH – ), Carbonate (CO 3<br />

–2 ) & Bicarbonate (HCO3 – ) ions)<br />

Test Results<br />

Calculation, Concentration mg/liter<br />

P f = 0 M f x 1220 = mg/L HCO 3<br />

– (Indicates Bicarbonate ion only)<br />

P f = M f P f x 340 = mg/L OH – (Indicates Hydroxide ion only)<br />

2P f < M f 2P f x 600 = mg/L CO 3<br />

–2 (Indicates Carbonate ion)<br />

(M f – 2P f ) x 1220 = mg/L HCO 3<br />

– (Indicates Bicarbonate ion)<br />

2P f = M f M f x 600 = mg/L CO 3<br />

–2 (Indicates Carbonate ion only)<br />

2P f > M f (2P f – M f ) x 340 = mg/L OH – (Indicates Hydroxide ion)<br />

(M f – P f ) x 1200 = mg/L CO 3<br />

–2 (Indicates Carbonate ion)<br />

<strong>OFI</strong>TE • 11302 Steeplecrest Dr. • Houston, TX 77065 USA • Phone (832) 320-7300 • http://www.ofite.com<br />

Version 1.1<br />

Date 03-04-03


<strong>OFI</strong> <strong>Testing</strong> <strong>Equipment</strong> – 146-00 <strong>Instructions</strong> – Workover and Completion Fluids Test Kit Page 3 of 17<br />

Procedure - Whole Mud Alkalinity, P m<br />

1. Measure 1 ml of drilling mud into the titration dish and dilute with 25 to 50 ml of distilled water.<br />

2. Add four or five drops of Phenolphthalein Indicator solution and while stirring, titrate with N/50 Sulfuric acid<br />

solution until the pink color just disappears. If the end point color change cannot be seen, it can be taken when<br />

the pH drops to pH 8.3 as measured on a pH meter. If cement contamination is suspected, the titration must be<br />

performed as rapidly as possible and the end-point is reported as the first disappearance of the pink color.<br />

3. Report the Phenolphthalein alkalinity of the whole mud, P m , as the number of milliliters of N/50 Sulfuric acid<br />

required per milliliter of mud.<br />

Procedure - Lime Content, Estimated<br />

1. Determine the P f and P m of the filtrate and whole mud as described in the Alkalinity test.<br />

2. Determine the volume fraction of water F w , using the percent of water from the liquid and solids as<br />

determined in a Retort analysis.<br />

F w = % Water by Volume<br />

100<br />

3. Report the Lime Content of the Fluid:<br />

Alternate Alkalinity Procedure<br />

Estimated Lime, lb/bbl = 0.26 (P m - F w x P f )<br />

Estimated Lime, kg/m 3 = 0.742 (P m - F w x P f )<br />

The P1/P2 Back Titration method is used to overcome some limitations of the P f /M f alkalinity method. A serious<br />

problem arises with anionic organic thinners, filtrate reducers and their degradation products which may contribute to a<br />

large portion of the alkalinity value, as well as masking the end point color change. These organic materials make a<br />

particularly large contribution to the M f alkalinity and thus render the test highly inaccurate in treated with organic<br />

thinners. If organic thinners are present in large amounts, the P1/P2 method should be used, but it also has its<br />

limitations. See the table below:<br />

Method Advantages Disadvantages<br />

P f /M f Traditional Method Interference with the M f titration<br />

2 Titrations, 1 Sample Bicarbonate result normally too high<br />

P1/P2 Eliminates interference 3 Titrations with 3 samples<br />

in M f titrations<br />

Caustic measurement critical<br />

Uses a toxic material - BaCl 2<br />

<strong>OFI</strong>TE • 11302 Steeplecrest Dr. • Houston, TX 77065 USA • Phone (832) 320-7300 • http://www.ofite.com<br />

Version 1.1<br />

Date 03-04-03


<strong>OFI</strong> <strong>Testing</strong> <strong>Equipment</strong> – 146-00 <strong>Instructions</strong> – Workover and Completion Fluids Test Kit Page 4 of 17<br />

<strong>Equipment</strong> - P1/P2 Alkalinity Method<br />

#147-16 Pocket pH Meter (Optional)<br />

#147-54 pH paper strips, range 7 - 14 pH<br />

#153-16 Graduate Cylinder, 25 ml x 2/10 ml, glass<br />

#153-18 Graduate Cylinder, 10 ml x 2/10 ml, glass<br />

#153-20 Graduate Cylinder, 5 ml x 1/10 ml, glass<br />

#153-26 Titration Dish, polyethylene<br />

#153-34 Pipette, 1 ml x 1/100 ml, glass<br />

#153-36 Pipette, 2 ml x 1/100 ml, glass<br />

#168-05 Stirring Rod, polyethylene<br />

Reagents:<br />

#206-02 Distilled Water, 16 oz<br />

#220-00 Phenolphthalein Indicator sol'n, 2 oz<br />

#260-01 Sodium Hydroxide sol'n, 16 oz<br />

#275-04 Hydrochloric Acid sol'n, 0.02N, 8 oz<br />

#285-07 Barium Chloride, 10 % sol'n, 8 oz<br />

Procedure - P1/P2 Alkalinity<br />

1. Determine the P f alkalinity as described above.<br />

2. Measure and add 1.0 ml of Filtrate to the titration dish, then add 25 ml of distilled water.<br />

3. Using a pipette add 2.0 ml of Sodium Hydroxide, 0.1N and stir well.<br />

4. Measure the pH of the solution with the pH paper or the pH meter.<br />

If the pH is 11.4 or greater proceed to the next step.<br />

If the pH is less than 11.4 add 2.0 ml more of 0.1N Sodium Hydroxide solution and then proceed. Exact<br />

measurements of Sodium Hydroxide is necessary to avoid serious errors.<br />

5. Using the 5 ml Graduate Cylinder, measure 3 ml of Barium Chloride and add to the titration dish.<br />

Caution: Do not use your mouth to pipette Barium Chloride as it is extremely poisonous.<br />

6. Add 2 to 4 drops of Phenolphthalein Indicator solution while stirring.<br />

7. Immediately titrate the mixture with the standard 0.02 normal Hydrochloric Acid to the first disappearance of<br />

the pink color - or to a pH of 8.3 with a pH meter. The color may reappear after a short time but do not continue<br />

the titration.<br />

8. Report the Alternate Alkalinity, P1, as the total ml of 0.02 normal Hydrochloric acid required to reach the<br />

phenolphthalein end point.<br />

9. Determine the Blank Alkalinity, P2. Omit the Filtrate, but repeat the above procedure from item 1 through 8 for<br />

determining the P1. Use exactly the same quantities of water and reagents in preparing the sample.<br />

10. Report the Blank Alkalinity, P2, as the total ml of 0.02 normal Hydrochloric acid required to titrate the<br />

reagent mixture to the phenolphthalein end point.<br />

<strong>OFI</strong>TE • 11302 Steeplecrest Dr. • Houston, TX 77065 USA • Phone (832) 320-7300 • http://www.ofite.com<br />

Version 1.1<br />

Date 03-04-03


<strong>OFI</strong> <strong>Testing</strong> <strong>Equipment</strong> – 146-00 <strong>Instructions</strong> – Workover and Completion Fluids Test Kit Page 5 of 17<br />

Calculation - P1/P2 Alkalinity Method<br />

When P1 > P2:<br />

When P1 < P2:<br />

OH – , mg/L = 340 (P1 – P2)<br />

CO 3<br />

–2 , mg/L = 1200 [Pf – (P1–P2)]<br />

HCO 3<br />

– , mg/L = 1220 (P2 – P1)<br />

CO 3<br />

-2 , mg/L = 1200 Pf<br />

WHOLE MUD ALKALINITY - OIL BASED DRILLING FLUID<br />

Procedure:<br />

1. Add 100 ml of Arcosolv PNP solvent to the 400 ml beaker<br />

2. Fill the 5 ml syringe with at least 3 ml of whole mud and discharge 2 ml into the beaker. Swirl the<br />

mixture until it is homogenous.<br />

3. Add 200 ml of distilled water.<br />

4. Add 15 drops of Phenolphthalein Indicator solution.<br />

5. While stirring with the magnetic stirrer, slowly titrate with N/10 Sulfuric Acid until the pink color just<br />

disappears. Continue stirring for one more minute, and if no pink color reappears, discontinue stirring. It may<br />

be necessary to stop the stirring to allow separation of the two phases in order to more clearly see the color in<br />

the aqueous phase.<br />

6. Let the sample stand for 5 minutes and if no pink color reappears, the Alkalinity end point has been reached. If<br />

the pink color returns, titrate a second time with the Sulfuric Acid solution. If the pink color still returns<br />

titrate a third time, but if the color returns after this third titration, call this the end point.<br />

7. Use the total ml of N/10 Sulfuric acid solution required to reach the end point to calculate the whole mud<br />

Alkalinity, V sa .<br />

Calculation:<br />

Alkalinity (Whole Mud), V sa =<br />

N/10 Sulfuric Acid, ml<br />

Mud Sample, ml<br />

<strong>OFI</strong>TE • 11302 Steeplecrest Dr. • Houston, TX 77065 USA • Phone (832) 320-7300 • http://www.ofite.com<br />

Version 1.1<br />

Date 03-04-03


<strong>OFI</strong> <strong>Testing</strong> <strong>Equipment</strong> – 146-00 <strong>Instructions</strong> – Workover and Completion Fluids Test Kit Page 6 of 17<br />

TOTAL HARDNESS TITRATION<br />

(Calcium and Magnesium)<br />

Kit Part No. 145-10<br />

INSTRUCTIONS:<br />

This test measures the amount of soluble calcium and magnesium ions in a mud filtrate. When measured<br />

simultaneously, the result is called total hardness. The calcium ion concentration can be determined separately and<br />

when it is subtracted from the total hardness, the magnesium ion concentration can then be determined. Different<br />

buffer-indicator combinations may be used for these determinations. It is imperative however not to intermix the<br />

various buffers and indicators.<br />

Procedure:<br />

Components:<br />

#153-26 Titration Dish, polyethylene<br />

#153-28 Stirring Rod, polyethylene<br />

#153-34 Pipette, glass, 1 ml x 1/100 ml<br />

#153-40 Pipette, glass, 10 ml x 1/10 ml<br />

Reagents:<br />

#205-02 Indicator sol’n, (Calmagite), Versenate Hardness, 2 oz<br />

#205-04 *Buffer sol’n., (Ammonium), 2 oz (UN2672)<br />

#205-06 Titration sol’n., (EDTA), 40 mg/L, 2 EPM, 8 oz<br />

#205-10 Titration sol’n., (EDTA), 400 mg/L, 20 EPM, 8 oz<br />

#206-02 Deionized Water, 16 oz<br />

• Hazardous Chemical – Must be specially packaged, and hazardous shipping charges will apply.<br />

1. Add 50 mls of deionized water to a titration dish.<br />

2. Add 20 to 40 drops (1 – 2 mls) of hardness buffer solution to the dish.<br />

3. Add 10 to 15 drops (0.5 – 0.75 mls) of hardness indicator solution to the mixture.<br />

4. If a wine-red to a purple color develops, the deionized water used in step 1, contains hardness. If there is a<br />

color change, add total hardness titrating solution drop wise while stirring until the water turns to a bright<br />

blue color.<br />

5. Using the pipette, add 1 ml of filtrate into the titration dish and stir. A wine red to purple color will again<br />

develop if calcium and/or magnesium is present in the filtrate solution.<br />

6. Add total hardness titrating solution, stirring continuously, until the sample again turns a bright blue color.<br />

Calculation:<br />

<strong>OFI</strong>TE • 11302 Steeplecrest Dr. • Houston, TX 77065 USA • Phone (832) 320-7300 • http://www.ofite.com<br />

Version 1.1<br />

Date 03-04-03


<strong>OFI</strong> <strong>Testing</strong> <strong>Equipment</strong> – 146-00 <strong>Instructions</strong> – Workover and Completion Fluids Test Kit Page 7 of 17<br />

epm Calcium and Magnesium = mls of Hardness Titrating solution used x 2<br />

(If the solution strength is 1 ml = 2 epm Hardness)<br />

epm Calcium and Magnesium = mls of Hardness Titrating solution used x 20<br />

(If the solution strength is 1 ml = 20 epm Hardness)<br />

Usually the magnesium ion content is negligible, so the total Hardness is reported as being all Calcium.<br />

Therefore:<br />

mg/L Calcium = epm Total Hardness x 20<br />

More or less sample can be used to obtain a titration between 1 and 10 mls, in which case the results obtained above<br />

should be divided by the ml of sample used.<br />

epm = equivalent parts per million<br />

Sometimes the test is erratic at low temperatures due to the change in solubility of the ingredients.<br />

If desired, separate test may be performed for Calcium and Magnesium individually using Calcium Buffer<br />

solution and Calver II Indicator powder. See separate instructions.<br />

Various Buffer Indicator Combinations:<br />

1. Versenate Buffer solution (#205-04, Ammonium Hydroxide, pH = 12 –13) with Versenate Indicator (#205-02,<br />

Calmagite).<br />

2. Manver Indicator (#205-27) with Ammonium Hydroxide Buffer (#205-04, pH = 12 –13)<br />

3. Eriochrome Black T indicator (#205-28) with Ammonium Hydroxide Buffer<br />

(#205-04 pH = 12 –13)<br />

Water Based Drilling Fluids<br />

All of these combinations may be used to determine Total Hardness.<br />

CALCIUM ANALYSIS<br />

INSTRUCTIONS<br />

When Hardness Titration solution (EDTA) is added to water or mud filtrate that contains both Calcium and Magnesium,<br />

the EDTA first combines with the Calcium. Calcium alone can be determined with EDTA when the pH of the sample is<br />

sufficiently high so that magnesium is precipitated out as the Hydroxide, and an indicator specific to Calcium is used.<br />

Several indicators will give color changes when all of the Calcium has been complexed by the EDTA at a pH in the<br />

range of 12 to 13. An end point obscured by dark organic components may be treated by oxidizing with Sodium<br />

Hypochlorite.<br />

<strong>Equipment</strong>:<br />

#153-14 Graduated Cylinder, 50 ml x 1.0 ml, glass<br />

#153-34 Pipette, 1.0 ml x 1/100 ml, glass<br />

<strong>OFI</strong>TE • 11302 Steeplecrest Dr. • Houston, TX 77065 USA • Phone (832) 320-7300 • http://www.ofite.com<br />

Version 1.1<br />

Date 03-04-03


<strong>OFI</strong> <strong>Testing</strong> <strong>Equipment</strong> – 146-00 <strong>Instructions</strong> – Workover and Completion Fluids Test Kit Page 8 of 17<br />

#153-36 Pipette, 2.0 ml x 1/10 ml, glass<br />

#153-38 Pipette, 5.0 ml x 1/10 ml, glass<br />

#153-40 Pipette, 10.0 ml x 1/10 ml, glass<br />

#153-51 Beaker, 250 ml, glass<br />

#168-01 Hot Plate, with Thermostat, 115 volt (Optional)<br />

(Required if the filtrate is colored)<br />

or<br />

#168-01-1 Hot Plate, with Thermostat, 230 volt (Optional)<br />

Reagents:<br />

#205-12 Titration sol'n (EDTA), Ver. Hard., 0.01M, 400 Mg/L, 16 oz<br />

#205-14 Calcium Buffer sol'n (NaOH), 1N, 2 oz<br />

#206-02 Deionized Water, 16 oz<br />

#210-00 Calver ® II Indicator Powder, 10 gram<br />

#230-25 Acetic Acid, Glacial, 8 oz (UN2789) (Optional)<br />

#261-00 Sodium Hypochlorite sol'n, 8 oz (UN1791) (Optional)<br />

#261-50 Masking Agent, 1:1:2 mixture by volume of<br />

Triethanolamine: Tetraethylenepentamine : Water, 16 oz<br />

(Optional)<br />

Procedure:<br />

WORK IN AN ADEQUATELY VENTILATED AREA.<br />

1. Using a pipette add at least 1.0 mL of sample to the Beaker. This volume will be used in the calculation<br />

formula below.<br />

If the filtrate is colorless or is only slightly colored, omit steps 2 - 5.<br />

2. Add 10.0 ml of Sodium Hypochlorite solution and mix.<br />

3. Add 1.0 ml of Glacial Acetic Acid and mix.<br />

4. Boil the sample for 5 minutes and maintain the sample volume by adding deionized water as required during<br />

boiling. Boiling removes excess Chlorine and this may be verified by immersing a strip of pH paper in the<br />

sample. If the paper is bleached white, continue boiling. A sufficiently boiled sample will have a pH of 5.0.<br />

5. Cool the sample and wash the sides of the beaker with deionized water.<br />

6. Using deionized water, dilute the sample to approximately 50 ml.<br />

7. Add 10 to 15 ml of Calcium Buffer solution, or sufficient Buffer solution to produce a pH of 12 - 13.<br />

Note: The presence of soluble Iron may interfere with the endpoint determination. If this is suspected, a mixture of<br />

Triethanolamine: Tetraethylenepentamine: Water (1:1:2 by volume) is a suitable masking agent.<br />

Add 1 ml of the mixture at this point.<br />

8. Add sufficient Calver ® II Indicator powder (0.1 to 0.2 grams) to produce a pink to wine-red color if<br />

Calcium is present. Too much indicator will obscure the endpoint. Note: Adding several drops of<br />

Methyl Orange Indicator solution may improve the visibility of the endpoint.<br />

9. While stirring, titrate with Hardness Titration solution (EDTA), to the end point. Calcium indicators produce a<br />

Red-to-Blue change and the endpoint is best described as that point where additional Titration solution will<br />

<strong>OFI</strong>TE • 11302 Steeplecrest Dr. • Houston, TX 77065 USA • Phone (832) 320-7300 • http://www.ofite.com<br />

Version 1.1<br />

Date 03-04-03


<strong>OFI</strong> <strong>Testing</strong> <strong>Equipment</strong> – 146-00 <strong>Instructions</strong> – Workover and Completion Fluids Test Kit Page 9 of 17<br />

produce no further Red-to-Blue color change. The Titration solution volume will be used in the calculation<br />

procedure.<br />

Calculation:<br />

Calcium, mg/L = 400 (Titration sol'n, ml3 )<br />

Sample Volume, ml 3<br />

Oil Based Drilling Fluids:<br />

<strong>Equipment</strong>:<br />

#153-53 Magnetic Stirrer, with Stirring Bar<br />

#153-64 Syringe, 5 ml, disposable,<br />

#154-75 Scoop, brass, 4 inch length<br />

#297-05 Mason Jar, with Lid, 16 oz<br />

Reagents:<br />

#205-17-3 Titration sol'n. (EDTA), 4000 mg/L Ca +2 , 0.1M, 200 EPM, 16 oz<br />

#206-04 Distilled Water, gal<br />

#210-00-1 Calver ® II Indicator powder, 100 gram<br />

#260-05 Sodium Hydroxide Buffer sol'n, 1.0 N, 8 oz (UN1824)<br />

#280-30 Arcosolv PNP, gal. (UN1993)<br />

Procedure:<br />

1. Add 100 mls of Arcosolv PNP solvent to a 16 oz jar<br />

2. Fill a new 5 ml syringe with whole mud past the 3 ml mark.<br />

3. Displace 2.0 mls of oil mud into the jar with the Arcosolv solution.<br />

4. Cap the jar and shake vigorously by hand for 1 minute.<br />

5. Add 200 mls of distilled water to the jar.<br />

6. Add *6.0 mls of Sodium Hydroxide Buffer sol'n, 1 N, to the mixture.<br />

7. Add *0.7 to 1.0 gram of Calver ® II Indicator powder. (Note: 5 scoops = 0.7 grams)<br />

Recap the jar tightly and shake vigorously again for 2 minutes. Set the jar aside for a few seconds to allow the<br />

upper and lower phases to separate. If a reddish color appears in the aqueous (lower) phase, this indicates that<br />

calcium is present.<br />

8. Agitate by placing the jar on a magnetic stirrer and drop in a stirring bar.<br />

9. Titrate very slowly, adding Titration solution (EDTA) drop by drop from a pipette, while stirring only fast<br />

enough to agitate the aqueous (lower) phase without remixing the upper and lower phases. A distinct color<br />

change from the reddish color to a blue-green occurs at the end point. Note the volume of Titration solution<br />

added.<br />

10. Use the mls of titration solution required to reach the end point to calculate the whole mud calcum.<br />

Calculation:<br />

Ca oil mud = Ca, mg = 4000 (0.1 M EDTA, mls)<br />

oil mud sample, L<br />

2.0 mls<br />

*Quantities have been revised since the last API issue - "R.P. 13B-2", Third Edition, Dated Feb<br />

Water Based Fluids<br />

CHLORIDE ION KIT<br />

PART NO. 144-40<br />

<strong>OFI</strong>TE • 11302 Steeplecrest Dr. • Houston, TX 77065 USA • Phone (832) 320-7300 • http://www.ofite.com<br />

Version 1.1<br />

Date 03-04-03


<strong>OFI</strong> <strong>Testing</strong> <strong>Equipment</strong> – 146-00 <strong>Instructions</strong> – Workover and Completion Fluids Test Kit Page 10 of 17<br />

This test measures the total soluble Chloride ion concentrations in the mud Filtrate. These Chlorides can come from<br />

Sodium Chloride, Calcium Chloride or Potassium Chloride. For the titration to work correctly, the pH of the filtrate<br />

needs to be slightly basic, around pH = 8.3, etc. Two chemical reactions take place simultaneously during the titration<br />

procedure.<br />

1. Ag + + Cl¯ ——————» AgCl<br />

2. 2Ag + + CrO 4<br />

–2 ————» Ag2 CrO 4<br />

The first reaction, the formation of Silver Chloride, accounts for the appearance of the white specs or milky appearance<br />

during the titration. The formation of Silver Chromate, which is red in color, will not start until all of the Chloride Ions<br />

are tied up as Silver Chloride. The Silver Nitrate will then react with the chromate from the Potassium Chromate<br />

Indicator to form Silver Chromate. So, for the above two reactions to occur, the filtrate needs to be weakly basic. (pH =<br />

8.3). High pH will precipitate out Silver Oxide.<br />

EQUIPMENT:<br />

#153-26 Titration Dish, polyethylene<br />

#153-28 Stirring Rod, polyethylene<br />

#153-34 Pipette, 1 ml x 1/100 ml, glass<br />

#153-40 Pipette, 10 ml x 1/10 ml, glass<br />

Reagents:<br />

#206-01 Distilled Water, 8 oz<br />

#215-00 *Potassium Chromate Indicator sol'n, 2 oz (UN3082)<br />

#220-00 Phenolphthalein Indicator sol'n, 2 oz<br />

#230-08 Sulfuric Acid, 0.02N (N/50), 8 oz<br />

#265-02 Silver Nitrate sol'n, 0.001G, 0.0282N, 16 oz<br />

#265-06 Silver Nitrate sol'n, 0.01G, 0.282N, 8 oz<br />

#285-00 Calcium Carbonate, precipitated. 35 grams<br />

Case:<br />

#144-35 Multi Kit Case, Diagonal design, Stainless Steel<br />

PROCEDURE - Water Based Muds:<br />

1. Pipette one or more milliliters (cm 3 ) of filtrate into the titration dish.<br />

2. Add 2 - 3 drops of Phenolphthalein Indicator sol'n to the filtrate.<br />

If a pink color appears - Titrate with N/50 Sulfuric acid until the color is discharged.<br />

If the filtrate is deeply colored - Add an additional 2 ml of N/50 Sulfuric Acid and stir.<br />

Then add 1 gram of Calcium Carbonate and stir.<br />

3. Add 25 - 50 ml of distilled water. If any water other than distilled water is used, the chlorides in it must be<br />

accounted for by titration before adding the filtrate to it.<br />

4. Add 5 - 10 drops of Potassium Chromate Indicator sol'n.<br />

5. Add Silver Nitrate sol'n drop-by-drop from the pipette while stirring continuously with the stirring rod, until the<br />

color changes from yellow to orange-red and persists for 30 seconds. Record the number of milliliters (cm 3 ) of<br />

Silver Nitrate sol'n required to reach the end point.<br />

A. If the chloride ion concentration of the filtrate exceeds 10,000 milli grams (mg/l) per liter use the<br />

<strong>OFI</strong>TE • 11302 Steeplecrest Dr. • Houston, TX 77065 USA • Phone (832) 320-7300 • http://www.ofite.com<br />

Version 1.1<br />

Date 03-04-03


<strong>OFI</strong> <strong>Testing</strong> <strong>Equipment</strong> – 146-00 <strong>Instructions</strong> – Workover and Completion Fluids Test Kit Page 11 of 17<br />

0.1 gram chloride-ion per cubic centimeter (0.282N) strength solution.<br />

B. If over 10 milliliters (cm 3 ) of Silver Nitrate is used, repeat the test with a smaller amount of filtrate.<br />

CALCULATION:<br />

To Convert Units:<br />

Chloride, mg/L =<br />

*1000 x mls of Silver Nitrate titrated)<br />

Filtrate Sample, mls<br />

Chloride, ppm = Chloride, mg/L<br />

Specific Gravity of Filtrate<br />

Salt (NaCl), mg/L = (1.65) Chloride, mg/L<br />

* Use 1000 as the multiplier for 0.0282N (.001gram) Chloride Ion solution, and 10,000 for the 0.282N (.01 gram) AgNO 3<br />

REMARKS:<br />

1. Avoid contact with Silver Nitrate and wash immediately with water if Silver Nitrate gets on skin or<br />

clothing.<br />

2. The end point of the reaction is when the Silver Chromate is first formed and it is reddish in color. When using<br />

the weak, or 0.001 gram sol'n, the end point is approached very gradually. Therefore the formation of the Silver<br />

Chromate can be seen by a color change from yellow to orange-red. If the strong (.01gram) solution is used the<br />

end point is reached much more rapidly and may go from yellow to red. As soon as the red color is seen, the titration<br />

procedure is complete.<br />

3. White lumps of Silver Chloride form when titrating high concentrations of salt. This should not be taken for<br />

the end point of the titration.<br />

Oil Based Fluids - Whole Mud Chloride<br />

The whole mud chloride test procedure for oil muds is a titration procedure which measures the volume of a standard<br />

Silver Nitrate required to react with the Chloride ions to form insoluble Silver Chloride salts. The Chloride value<br />

reported for the whole mud may be assigned to the water phase of the solution up to the saturation point. The water<br />

soluble salt concentration indicates how effective an oil mud is in controlling shale through the "aqueous phase activity"<br />

concept. The water phase salinity value is also needed to adjust the retort value to obtain a corrected solids content for<br />

the oil mud. The test procedure uses the same sample as used in the whole mud Alkalinity test provided the sample is<br />

acidic (pH < 7).<br />

<strong>Equipment</strong> - Oil Based Fluids:<br />

#153-29-1 Syringe, glass tip, 5 ml<br />

#153-51-1 Beaker, 400 ml, glass<br />

#153-53 Magnetic Stirrer, with Stir Bars<br />

<strong>OFI</strong>TE • 11302 Steeplecrest Dr. • Houston, TX 77065 USA • Phone (832) 320-7300 • http://www.ofite.com<br />

Version 1.1<br />

Date 03-04-03


<strong>OFI</strong> <strong>Testing</strong> <strong>Equipment</strong> – 146-00 <strong>Instructions</strong> – Workover and Completion Fluids Test Kit Page 12 of 17<br />

Reagents:<br />

#206-02 Distilled Water, 16 oz<br />

#215-02 Potassium Chromate sol'n, 8 oz (UN3082)<br />

#220-01 Phenolphthalein Indicator sol'n, 8 oz<br />

#230-10 Sulfuric Acid sol'n, N/10, 8 oz<br />

#265-08 Silver Nitrate sol'n, 0.01g, 0.282N, 16 oz<br />

#280-30 Arcosolv PNP, gal.<br />

Procedure - Oil Based Fluids<br />

Perform the whole mud alkalinity test first as follows:<br />

1. Add 100 ml of Arcosolv PNP solvent to the 400 ml beaker.<br />

2. Fill the 5 ml syringe with at least 3 mls of whole mud and discharge 2 ml into the beaker. Swirl the<br />

mixture until it is homogenous.<br />

3. Add 200 ml of distilled water.<br />

4. Add 15 drops of Phenolphthalein Indicator solution.<br />

5. While stirring with the magnetic stirrer, slowly titrate with N/10 Sulfuric Acid until the pink color just<br />

disappears. Continue stirring for one more minute, and if no pink color reappears, discontinue stirring. It may<br />

be necessary to stop the stirring to allow separation of the two phases in order to more clearly see the color in<br />

the aqueous phase.<br />

6. Let the sample stand for 5 minutes and if no pink color reappears, the Alkalinity end point has been reached. If<br />

thepink color returns, titrate a second time with the Sulfuric Acid solution. If the pink color still returns titrate a<br />

third time, but if the color returns after this third titration, call this the end point.<br />

7. After running the Alkalinity test make sure the mixture to be titrated for Chlorides is acidic (pH < 7) by<br />

adding 10 - 20 drops of N/10 Sulfuric Acid.<br />

8. Add *3.0 mls of Potassium Chromate Indicator solution.<br />

9. While stirring with a Magnetic Stirrer, slowly titrate with 0.282N Silver Nitrate solution until a salmon pink<br />

color remains stable for at least one minute. It may be necessary to stop the stirring of the mixture and allow<br />

separation of the two phases to occur in order to more clearly see the color in the aqueous phase.<br />

10. Note the total milliliters of 0.282N Silver Nitrate required to reach the end point to calculate the whole mud<br />

Chloride.<br />

Calculation:<br />

Chloride, Whole Mud (Cl om ) Mg/l = 10,000 ( ml Silver Nitrate titrated)<br />

ml of sample<br />

*Quantity has been revised since the last API issue - "R.P. 13B-2", Third Edition, Dated Feb. 1998.<br />

<strong>OFI</strong>TE • 11302 Steeplecrest Dr. • Houston, TX 77065 USA • Phone (832) 320-7300 • http://www.ofite.com<br />

Version 1.1<br />

Date 03-04-03


<strong>OFI</strong> <strong>Testing</strong> <strong>Equipment</strong> – 146-00 <strong>Instructions</strong> – Workover and Completion Fluids Test Kit Page 13 of 17<br />

CALCIUM SULFATE ANALYSIS<br />

INSTRUCTIONS<br />

The Calcium Sulfate content of a drilling fluid is determined by using EDTA Titration solution to determine the total<br />

Calcium in both mud filtrate and in whole mud. The total and undissolved Calcium Sulfate content of the drilling fluid<br />

can then be calculated.<br />

<strong>Equipment</strong>:<br />

#147-53 pH paper, range 1 - 14<br />

#153-14 Cylinder, graduated, 50 ml x 1 ml, glass<br />

#153-26 Titration Dish, Polyethylene<br />

#153-34 Pipette, 1 ml x 1/100 ml, glass<br />

#153-36 Pipette, 2 ml x 1/100 ml, glass<br />

#153-38 Pipette, 5 ml x 1/10 ml, glass<br />

#153-40 Pipette, 10 ml x 1/10 ml, glass<br />

#168-01 Hot Plate, 115 volt (Required if filtrate is colored)<br />

Reagents:<br />

#205-12 Tittration solution, EDTA, 400 mg/L Ca +2 , 20 EPM, 16 oz<br />

#206-02 Distilled Water, 16 oz<br />

#210-00 Calver ® II Indicator powder, 10 gram<br />

#230-20 *Glacial Acetic Acid, 8 oz (UN2789)<br />

#260-05 *Sodium Hydroxide, 1N, 8 oz (UN1824)<br />

#261-00 Sodium Hypochlorite sol'n, 8 oz (UN1791)<br />

#261-50 Triethanolamine masking agent, 16 oz<br />

Note: Also necessary will be the following items:<br />

#140-30 Filter Press, API, full area, with CO 2 Pressuring assembly<br />

#165-00-1 Retort Kit, 10 ml, with stainless steel case, 115 volt<br />

or<br />

#165-10-1 Retort Kit, 10 ml, with stainless steel case, 220 volt<br />

Procedure:<br />

1. Add 5 ml of whole mud to 245 ml of distilled water.<br />

2. Stir the mixture for 15 minutes and filter the diluted mixture through a standard API Filter Press, collecting only<br />

clear filtrate. Also filter the undiluted whole mud through the standard API Filter Press.<br />

3. Using the 10 ml pipette, add 10 ml of the clear filtrate to the titration dish. Titrate with EDTA titration solution<br />

to the calcium endpoint, and call this volume EDTA V t. The procedure is as follows:<br />

<strong>OFI</strong>TE • 11302 Steeplecrest Dr. • Houston, TX 77065 USA • Phone (832) 320-7300 • http://www.ofite.com<br />

Version 1.1<br />

Date 03-04-03


<strong>OFI</strong> <strong>Testing</strong> <strong>Equipment</strong> – 146-00 <strong>Instructions</strong> – Workover and Completion Fluids Test Kit Page 14 of 17<br />

A. Add at least 1.0 mL of sample to the Beaker.<br />

If the filtrate is colorless or is only slightly colored, omit steps B - E.<br />

B. Add 10.0 ml of Sodium Hypochlorite solution and mix.<br />

C. Add 1.0 ml of Glacial Acetic Acid and mix.<br />

D. Boil the sample for 5 minutes and maintain the sample volume by adding distilled water as required<br />

during boiling. Boiling removes excess Chlorine and this may be verified by immersing a strip of Ph<br />

paper in the sample. If the paper is bleached white, continue boiling. A sufficiently boiled sample will<br />

have a pH of 5.0.<br />

E. Cool the sample and wash the sides of the beaker with distilled water.<br />

F. With distilled water, dilute the sample to approximately 50 ml.<br />

G. Add 10 to 15 ml of Calcium Buffer solution, or sufficient Buffer solution to produce a pH of 12 - 13.<br />

H. Optional: The presence of soluble Iron may interfere with the endpoint determination. If this is<br />

suspected, a mixture of Triethanolamine: Tetraethylenepentamine : Water (1:1:2 by volume) is a<br />

suitable masking agent.<br />

I. Add sufficient Calver ® II Indicator powder (0.1 to 0.2 grams) to produce a pink to wine-red color if<br />

Calcium is present. Too much indicator will obscure the endpoint. Note: Adding several drops of<br />

Methyl Orange Indicator solution may improve the visibility of the endpoint.<br />

J. While stirring, titrate with Hardness Titration solution (EDTA), to the end point. Calcium indicators<br />

produce a Red-to-Blue change and the endpoint is best described as that point where additional Titration<br />

solution will produce no further Red-to-Blue color change. The Titration solution volume will be used<br />

in the calculation procedure.<br />

3. Using the 1 ml pipette, add 1.0 ml of the original undiluted mud filtrate to the titration dish. Titrate with<br />

EDTA titration solution to the calcium endpoint, as above. Call this volume EDTA V f .<br />

4. Retort the whole mud. Determine the volume fraction of water in the mud by using the volume percent of water<br />

from the liquid and solids determination based upon the following equation.<br />

Volume Fraction of Water, F w =<br />

Volume Percent Water<br />

100<br />

Calculation:<br />

Total Calcium Sulfate, lb/bbl = 2.38 ( EDTA V t )<br />

The excess undissolved Calcium Sulfate content of the mud in pounds per barrel may be calculated as follows:<br />

Excess Calcium Sulfate, lb/bbl = 2.38 (EDTA V t) – 0.48 [EDTA V f x (F w )]<br />

Procedure:<br />

MAGNESIUM CONTENT<br />

The Hardness Titration test and the Calcium Titration test must be performed first.<br />

The total milliliters of Versenate Titration solution (EDTA), 1 mL = 400 mg/L Ca ++ used in the determination of total<br />

hardness, minus the milliliters of Versenate Titration solution (EDTA) used in the determination of calcium ion, times<br />

243 equals the mg/L of Magnesium.<br />

<strong>OFI</strong>TE • 11302 Steeplecrest Dr. • Houston, TX 77065 USA • Phone (832) 320-7300 • http://www.ofite.com<br />

Version 1.1<br />

Date 03-04-03


<strong>OFI</strong> <strong>Testing</strong> <strong>Equipment</strong> – 146-00 <strong>Instructions</strong> – Workover and Completion Fluids Test Kit Page 15 of 17<br />

Calculation:<br />

Mg ++ , mg/L =<br />

(mL EDTA, Total Hardness – mL EDTA, Ca++) x 243<br />

mL of Filtrate<br />

POTASSIUM & POTASSIUM CHLORIDE<br />

<strong>OFI</strong> Part No. 285-09<br />

Centrifuge Method: > 5000 Milligrams per Liter KCl (1% KCl)<br />

Potassium ion is used in drilling fluids to aid in the stabilization of shales and to control swelling clays. The accurate<br />

determination of the potassium ion content is necessary to control the properties of the drilling fluid. This procedure is<br />

used to measure the potassium ion content in mud filtrates at levels above 5000 milligrams per liter or 3.5 pounds per<br />

barrel KCl. The procedure is most accurate in the range between 10,000 mg/L (2%) and 40,000 mg/L (8%) KCl and is<br />

suspect below 10,000 mg/L (2%) KCl. Potassium is precipitated in a centrifuge tube as the perchlorate salt, and then the<br />

precipitate volume is measured. The potassium ion content is then read from a prepared standard curve.<br />

<strong>Equipment</strong>:<br />

#153-21 Centrifuge Tube, Kolmer, 10 ml (2 supplied)<br />

#153-25-2 *Centrifuge, Hand Crank, with dual head and shields<br />

#153-38 Pipette, 5 ml x 1/10 ml, glass<br />

Reagents:<br />

#206-02 Distilled Water, 16 oz<br />

#285-10 *Sodium Perchlorate sol'n, 16 oz (UN1502)<br />

#285-11 Potassium Chloride sol'n (standard), 4 oz<br />

Procedure - Generating a Standard Calibration Curve:<br />

1. A standard calibration curve is required for each type of centrifuge. A minimum of three points or test<br />

procedures (3.5, 10.5 & 17.5 pounds per barrel) is required to obtain an accurate graph.<br />

2. Samples can be prepared by using the standard potassium chloride solution where 0.5 ml of standard<br />

potassium chloride solution is equivalent to 3.5 pounds per barrel KCl.<br />

3.5 ppb KCl = 1 % KCL = 5000 mg/L ≅ 0.5 ml standard KCl solution.<br />

Therefore -<br />

A. To obtain 3.5 ppb equivalent KCl or 1% KCl use 0.5 ml of standard potassium chloride.<br />

B. To obtain 10.5 ppb equivalent KCl or 3% KCl use 1.5 ml of standard potassium chloride.<br />

C. To obtain 17.5 ppb equivalent KCl or 5% KCl use 2.5 ml of standard potassium chloride<br />

3. Dilute the samples in the centrifuge tube to the 7.0 ml mark with distilled water and agitate the mixture.<br />

4. Add 3 ml of standard sodium perchlorate solution to each centrifuge tube and do not agitate.<br />

5. Centrifuge at a constant speed of approximately 1800 revolutions per minute for one minute and read the<br />

precipitate volume immediately. It will be necessary to counterbalance the centrifuge tube with another tube<br />

and 10 ml of liquid of the same weight.<br />

A fairly constant 1800 revolutions per minute can be obtained by determining the number of revolutions of the rotor per<br />

each turn of the crank. Move the crank very slowly and count the number of revolutions of the rotor head during one turn<br />

of the crank. For example, 15 revolutions of the rotor per one turn of the crank. Calculate the number of crank turns<br />

required to obtain 1800 revolutions of the rotor head. In the example to obtain 1800 revolutions of the head would require<br />

120 turns of the crank (1800/15). Thus the crank must be turned 120 times in one minute to obtain the rate of 1800<br />

<strong>OFI</strong>TE • 11302 Steeplecrest Dr. • Houston, TX 77065 USA • Phone (832) 320-7300 • http://www.ofite.com<br />

Version 1.1<br />

Date 03-04-03


<strong>OFI</strong> <strong>Testing</strong> <strong>Equipment</strong> – 146-00 <strong>Instructions</strong> – Workover and Completion Fluids Test Kit Page 16 of 17<br />

revolutions per minute. At this rate in 5 seconds the handle must be turned 10 times (120/60) (5). By counting the crank<br />

turns in 5 seconds and adjusting the rate to obtain the required number of turns, a constant 1800 revolutions per minute<br />

should be obtained in 15 to 20 seconds. The interval used to adjust to the 1800 revolutions per minute should be included<br />

in the centrifuge time of the sample.<br />

6. Clean the centrifuge tube immediately after use to facilitate ease of cleaning.<br />

7. Plot the volume of precipitate in milliliters versus pounds per barrel of KCl or milliliters of precipitate versus<br />

percent (%) potassium chloride using rectangular graph paper as shown below.<br />

Procedure - Sample <strong>Testing</strong><br />

1. Measure the appropriate volume of filtrate into the centrifuge tube. See the table below for the volume:<br />

Filtrate Volumes to Be Used At Various KCL Concentrations<br />

Concentration<br />

Filtrate Volume<br />

Range KCl K + to use<br />

(lb/bbl (mg/L) (ml)<br />

3.5 - 18 5,250 - 27,000 7.0<br />

18 - 35 27,000 - 52,500 3.5<br />

35 - 70 52,500 - 105,000 2.0<br />

over 70 over 105,000 1.0<br />

2. Dilute to 7.0 ml if necessary with distilled water in the tube and agitate the mixture.<br />

3. Add 3.0 ml of standard sodium perchlorate solution and do not agitate. If potassium is present precipitation<br />

occurs at once.<br />

4. Centrifuge at a constant speed of approximately 1800 rpm for one minute. Read the precipitate volume<br />

immediately and then record it.<br />

5. Add 2 to 3 drops of the sodium perchlorate solution to the tube. If a precipitate still forms, the total<br />

amount of potassium has not been measured. Refer to the table above and use the next smaller filtrate<br />

volume. Repeat the procedure in items 1 - 5 above.<br />

6. Determine the potassium chloride concentration by comparing the precipitate volume measured with the<br />

standard calibration curve as prepared above. Report the potassium concentration as pounds per barrel KCl.<br />

The potassium concentration may also be reported as milligrams per liter potassium ion. If the filtrate potassium<br />

chloride concentration from the standard calibration curve exceeds an 18 pounds per barrel reading, accuracy of<br />

the results is reduced. For more accurate results, use the next smaller filtrate volume as noted in the table above,<br />

and repeat the steps 1 through 5 above.<br />

Calculation:<br />

KCl in filtrate, lb/bbl = 7 ( Value from standard curve)<br />

<strong>OFI</strong>TE • 11302 Steeplecrest Dr. • Houston, TX 77065 USA • Phone (832) 320-7300 • http://www.ofite.com<br />

Version 1.1<br />

Date 03-04-03


<strong>OFI</strong> <strong>Testing</strong> <strong>Equipment</strong> – 146-00 <strong>Instructions</strong> – Workover and Completion Fluids Test Kit Page 17 of 17<br />

filtrate used, ml<br />

K + in filtrate, mg/L = 1500 (KCl in filtrate, lb/bbl)<br />

High Potassium Chloride Concentrations<br />

If the potassium concentration of the filtrate is greater than 35,000 mg/L (7%) KCl, dilution may be accomplished as<br />

follows:<br />

1. Pipette 2.0, 3.0 or 4.0 ml of filtrate into the centrifuge tube and dilute with distilled water to the 7.0 ml mark.<br />

2. Correct the results by multiplying % KCl by 7, divided by the volume of filtrate used (2.0, 3.0 or 4.0 ml)<br />

Remarks<br />

1. The potassium ion test must be made with the same centrifuge used to construct the standard curve. Turn the<br />

centrifuge at a constant speed of 1800 rpm.<br />

2. Sodium and potassium perchlorates are explosive in the dry state if heated to a high temperature or allowed to<br />

contact organic reducing agents. The perchlorates are not hazardous if kept water wet. They will decompose<br />

harmlessly if dispersed in a bucket of water.<br />

3. Salts, lignosulfonates, polymers, clay and pH do not interfere with the test.<br />

4. The precipitate may become slightly discolored in high lignosulfonate concentrations, but the test is not<br />

adversely affected.<br />

5. Only the Kolmer-type clinical centrifuge tubes are satisfactory for the test.<br />

6. The calibration curve does not need to be remade for every potassium determination, however the validity of the<br />

curve should be rechecked regularly by re-running at least one standard.<br />

7. A new standard graph must be prepared if a different model of centrifuge is used.<br />

<strong>OFI</strong>TE • 11302 Steeplecrest Dr. • Houston, TX 77065 USA • Phone (832) 320-7300 • http://www.ofite.com<br />

Version 1.1<br />

Date 03-04-03

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!