Microplate Barcode Labeler User Guide - Agilent Technologies
Microplate Barcode Labeler User Guide - Agilent Technologies
Microplate Barcode Labeler User Guide - Agilent Technologies
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Chapter 4: Setting up the VCode<br />
VCode <strong>User</strong> <strong>Guide</strong><br />
73<br />
Method two<br />
To set the robot pickup angle by manually moving the plate<br />
stage:<br />
1. In the Device Selection group box of the Print Job page of<br />
PlateTag, make sure that an Ethernet-based profile is selected if you<br />
are connecting using Ethernet or a serial-based profile is selected if<br />
you are connecting using serial.<br />
2. In PlateTag, select View > Diagnostics.<br />
The plate stage moves and the VCode Diagnostics dialog box opens<br />
after a few seconds.<br />
3. If the applicator head is pointing towards the plate stage:<br />
a. Click the I/O functions tab.<br />
b. Click Rotate to printer.<br />
The head rotates towards the printer, moving it out of the plate<br />
stage rotation path.<br />
c. Click the Profiles tab.<br />
!! DAMAGE HAZARD !! Before moving the stage, make sure<br />
that the head is not in the way.<br />
4. Click the Stage Setup tab.<br />
5. Make sure that Disable stage auto-repositioning is selected.<br />
6. Manually rotate the plate stage so that the east side of the plate faces<br />
the applicator head.<br />
7. Click the I/O functions tab and then click Rotate to Stage.<br />
8. Click Extend Head.<br />
9. Fine-tune the angle so that the applicator head is lined up with the<br />
plate.<br />
10. Click Teach here.<br />
11. Click Retract head.<br />
12. Click Rotate to Printer.<br />
Related information<br />
For more information about...<br />
Stage auto-repositioning<br />
Manually rotating the stage<br />
See...<br />
“About plate stage auto-repositioning”<br />
on page 68<br />
“Rotating the plate stage manually” on<br />
page 70