Chapter 3: Installing the VCode<br />
VCode <strong>User</strong> <strong>Guide</strong><br />
31<br />
Setting the air pressure<br />
Introduction<br />
After connecting the air tubing to the VCode you set the air pressure.<br />
Procedure<br />
To set the air pressure:<br />
1. Turn on the air supply to the VCode.<br />
Typically, the mechanism for turning on the air is a pressure<br />
regulator located on the wall of your lab.<br />
2. Adjust the air pressure using the pressure regulator knob at the back<br />
of the VCode:<br />
a. Pull out the knob until you feel a click.<br />
b. Turn the knob to increase or decrease pressure, while watching<br />
the pressure gauge.<br />
The pressure should be set to 68–72 psi (0.47–0.50 MPa).<br />
Knob<br />
c. Push the knob back in.<br />
Related information<br />
For more information<br />
about...<br />
Connecting the air tubing<br />
See...<br />
❑ “Connecting the air tubing: North<br />
America” on page 29<br />
❑ “Connecting the air tubing: Europe” on<br />
page 30