Chapter 3: Installing the VCode<br />

VCode <strong>User</strong> <strong>Guide</strong><br />

31<br />

Setting the air pressure<br />

Introduction<br />

After connecting the air tubing to the VCode you set the air pressure.<br />

Procedure<br />

To set the air pressure:<br />

1. Turn on the air supply to the VCode.<br />

Typically, the mechanism for turning on the air is a pressure<br />

regulator located on the wall of your lab.<br />

2. Adjust the air pressure using the pressure regulator knob at the back<br />

of the VCode:<br />

a. Pull out the knob until you feel a click.<br />

b. Turn the knob to increase or decrease pressure, while watching<br />

the pressure gauge.<br />

The pressure should be set to 68–72 psi (0.47–0.50 MPa).<br />

Knob<br />

c. Push the knob back in.<br />

Related information<br />

For more information<br />

about...<br />

Connecting the air tubing<br />

See...<br />

❑ “Connecting the air tubing: North<br />

America” on page 29<br />

❑ “Connecting the air tubing: Europe” on<br />

page 30