Chapter 3: Installing the VCode<br />

VCode <strong>User</strong> <strong>Guide</strong><br />

29<br />

Connecting the air tubing: North America<br />

Introduction<br />

This topic is for customers in North America only. It explains how to<br />

connect the air tubing to the VCode.<br />

Warnings<br />

!! INJURY HAZARD !! Switch off the air supply before installing<br />

your VCode. When air is flowing, air tubing can injure you if it<br />

whips around and strikes you. If you have not been trained in the<br />

correct way to handle air tubing, contact your facilities department<br />

or the Velocity11 Service Center.<br />

!! DAMAGE HAZARD !! Make sure that the air tubing coming<br />

into your VCode is properly filtered for moisture. Moist air might<br />

shorten the life of your VCode. If the quality of your air supply is in<br />

question, please contact Velocity11®.<br />

Connection<br />

procedure<br />

The following procedure is used to set up a direct connection. If you<br />

have another instrument connected to the house air supply, you can use<br />

the supplied quick-release T-connector to connect to the existing air<br />

tubing.<br />

To install the air tubing:<br />

1. Cut the supplied 1/4-inch tubing to the length you need.<br />

2. If your house air uses a threaded connection at the supply end,<br />

attach the 1/4-inch NPT fitting to the connection.<br />

3. Plug one end of the tubing into your lab’s air supply.<br />

4. Plug the other end of the 1/4-inch tubing into the orange quickrelease<br />

air supply port at the back of the VCode.<br />

00150c<br />

Vcode<br />

air hose<br />

00150c<br />

5. Test the installation by gently tugging the air tubing.<br />

If you feel resistance, the line has been properly installed.