28<br />
Chapter 3: Installing the VCode<br />
VCode <strong>User</strong> <strong>Guide</strong><br />
A 1/4” nipple (white plastic tube)<br />
A quick-release T-hose fitting, 6 mm<br />
❑ A VCode software and documentation CD-ROM<br />
❑ A printed VCode <strong>User</strong> <strong>Guide</strong><br />
Air fittings<br />
Depending on the configuration of your air supply system, you may<br />
need to use one or more of the air fittings supplied in the box. If your air<br />
system has an unusual configuration, you will need to supply your own<br />
fittings.<br />
Computer requirements<br />
Requirements<br />
Before starting the installation, make sure that your computer meets the<br />
following requirements:<br />
❑ A computer that is running the Microsoft ® Windows ® NT 4.0, 2000,<br />
or XP operating system<br />
❑ Either:<br />
A serial DB9 RS-232 port<br />
<br />
An available Ethernet network card<br />
Lab requirements<br />
Requirements<br />
Before installing the VCode, make sure that your laboratory has:<br />
❑ A supply of clean, dry compressed air: 2.5 cfm at 80 psi (5.3 L/s at<br />
0.55 MPa)<br />
❑ A power supply meeting one of the following specifications:<br />
North American: 120 VAC, 50–60 Hz, 2 A<br />
European: 20–240 VAC, 50 Hz, 2 A<br />
❑ Sufficient bench space: 28.5 cm x 61.0 cm (11.2 inches x<br />
24.0 inches) (not including the computer)