Chapter 3: Installing the VCode<br />

VCode <strong>User</strong> <strong>Guide</strong><br />

A 1/4” nipple (white plastic tube)<br />

A quick-release T-hose fitting, 6 mm<br />

❑ A VCode software and documentation CD-ROM<br />

❑ A printed VCode <strong>User</strong> <strong>Guide</strong><br />

Air fittings<br />

Depending on the configuration of your air supply system, you may<br />

need to use one or more of the air fittings supplied in the box. If your air<br />

system has an unusual configuration, you will need to supply your own<br />

fittings.<br />

Computer requirements<br />

Requirements<br />

Before starting the installation, make sure that your computer meets the<br />

following requirements:<br />

❑ A computer that is running the Microsoft ® Windows ® NT 4.0, 2000,<br />

or XP operating system<br />

❑ Either:<br />

A serial DB9 RS-232 port<br />

<br />

An available Ethernet network card<br />

Lab requirements<br />

Requirements<br />

Before installing the VCode, make sure that your laboratory has:<br />

❑ A supply of clean, dry compressed air: 2.5 cfm at 80 psi (5.3 L/s at<br />

0.55 MPa)<br />

❑ A power supply meeting one of the following specifications:<br />

North American: 120 VAC, 50–60 Hz, 2 A<br />

European: 20–240 VAC, 50 Hz, 2 A<br />

❑ Sufficient bench space: 28.5 cm x 61.0 cm (11.2 inches x<br />

24.0 inches) (not including the computer)