Krohne DK32 and DH34 Variable Area Flow Meters ... - Instrumart
Krohne DK32 and DH34 Variable Area Flow Meters ... - Instrumart Krohne DK32 and DH34 Variable Area Flow Meters ... - Instrumart
4 ELECTRICAL CONNECTIONS DK32 - DK34 - DK37 Connection three-wire reed contact Figure 4-2: Electrical connection of reed contact limit switch Strand colours for flowmeters with preassembled cables: 1 Silicone-insulated wire - yellow/green / FEP-insulated wire - red 2 Silicone-insulated wire - brown / FEP-insulated wire - brown 3 Silicone-insulated wire - blue / FEP-insulated wire - blue 4.3 Setting the limit switch Set the pointers to the desired limit values as a min. contact 1 and max. contact 2 using a slip coupling along the scale. Figure 4-3: Limit switch settings Adjusting the limit switch with reed contact: • Slacken nut 3 • Set reed cartridge to the required value • Secure with nut 3 Comment The reed contact is actuated directly by the float magnet. The desired operating point can only be determined in measuring mode. A reference to the scale and/or pointer cannot be established. 16 www.krohne.com 01/2012 - 4000287103 MA DK32-34-37-R05-en
DK32 - DK34 - DK37 ELECTRICAL CONNECTIONS 4 4.4 DK37/M8M limit switches The limit switches can be set over the entire measuring range using the maximum pointer. The set limit values are displayed on the scale. The pointers are set to the desired limit values using a slip coupling along the scale. Figure 4-4: Limit switch setting 1 Maximum pointer, switching point indicator 2 Limit switch 3 Connection board 4 Connection terminal 01/2012 - 4000287103 MA DK32-34-37-R05-en www.krohne.com 17
- Page 1 and 2: DK32 - DK34 - DK37 Handbook Variabl
- Page 3 and 4: DK32 - DK34 - DK37 CONTENTS 1 Safet
- Page 5 and 6: DK32 - DK34 - DK37 SAFETY INSTRUCTI
- Page 7 and 8: DK32 - DK34 - DK37 SAFETY INSTRUCTI
- Page 9 and 10: DK32 - DK34 - DK37 DEVICE DESCRIPTI
- Page 11 and 12: DK32 - DK34 - DK37 DEVICE DESCRIPTI
- Page 13 and 14: DK32 - DK34 - DK37 INSTALLATION 3 3
- Page 15: DK32 - DK34 - DK37 ELECTRICAL CONNE
- Page 19 and 20: DK32 - DK34 - DK37 ELECTRICAL CONNE
- Page 21 and 22: DK32 - DK34 - DK37 ELECTRICAL CONNE
- Page 23 and 24: DK32 - DK34 - DK37 SERVICE 6 6.1 Ma
- Page 25 and 26: DK32 - DK34 - DK37 SERVICE 6 6.4.2
- Page 27 and 28: DK32 - DK34 - DK37 TECHNICAL DATA 7
- Page 29 and 30: DK32 - DK34 - DK37 TECHNICAL DATA 7
- Page 31 and 32: DK32 - DK34 - DK37 TECHNICAL DATA 7
- Page 33 and 34: DK32 - DK34 - DK37 TECHNICAL DATA 7
- Page 35 and 36: DK32 - DK34 - DK37 TECHNICAL DATA 7
- Page 37 and 38: DK32 - DK34 - DK37 TECHNICAL DATA 7
- Page 39 and 40: DK32 - DK34 - DK37 NOTES 8 01/2012
<strong>DK32</strong> - DK34 - DK37<br />
ELECTRICAL CONNECTIONS 4<br />
4.4 DK37/M8M limit switches<br />
The limit switches can be set over the entire measuring range using the maximum pointer. The<br />
set limit values are displayed on the scale. The pointers are set to the desired limit values using<br />
a slip coupling along the scale.<br />
Figure 4-4: Limit switch setting<br />
1 Maximum pointer, switching point indicator<br />
2 Limit switch<br />
3 Connection board<br />
4 Connection terminal<br />
01/2012 - 4000287103 MA <strong>DK32</strong>-34-37-R05-en<br />
www.krohne.com<br />
17