Development and validation of stability indicating method for ...

Development and validation of stability indicating method for ... Development and validation of stability indicating method for ...

journal.chemistrycentral.com
from journal.chemistrycentral.com More from this publisher
30.03.2014 Views

Walash et al. Chemistry Central Journal 2011, 5:61 http://journal.chemistrycentral.com/content/5/1/61 Page 4 of 11 Fluorescence Intensity 160 140 120 100 80 60 40 20 a 0 250 300 350 400 450 500 550 600 650 700 750 800 a) Excitation Spectrum of drug. b) Emission Spectrum of drug. a`) Excitation Spectrum of Blank. b`) Emission Spectrum of Blank. a` Wavelength (nm.) Figure 3 Fluorescence spectra of the reaction product. (A,B) SER ( 2.5 μg/mL) with 0.2% NBD-Cl at pH 7.8. (A’, B’) blank with 0.2% NBD-Cl at pH 7.8. was obtained at pH 7.7 and remained constant up to 7.9 after which the absorbance of the reaction product began to decrease gradually until pH 9. Therefore, pH of 7.8 ± 0.1 was chosen as the optimum one (Figure 4). Other buffers having the same pH value such as phosphate and hexamine were tried and compared with 0.2 M borate buffer. Borate buffer was found to be superior to other buffers having the same pH value since the net fluorescence intensity was highest in case of borate buffer. This is probably, because the rate of hydrolysis of NBD-Cl to NBD-OH was much slower. This result is in agreement with that of Miyano et al [24]. Effect of concentration of NBD-Cl The influence of the concentration of NBD-Cl was studied using different volumes of 0.2% w/v solution of the reagent. It was found that, the reaction of NBD-Cl with b b` SER started upon using 0.2 mL of the reagent in the presence of borate buffer (pH 7.8). Increasing the volume of the reagent, produces a proportional increase in the absorbance of the reaction product up to 0.8 mL and remains constant up to 1.2, after which further increase produces a gradual decrease in the absorbance value. Therefore, 1 mL of 0.2% w/v of NBD-Cl solution was chosen as the optimal volume of the reagent (Figure 5). Effect of heating temperature Different temperature settings were used with constant heating time. Increasing the temperature of the water bath was found to produce a proportional increase in the absorbance of the reaction product up to 55°C and remained constant until 65°C after which further increase in the temperature produces a gradual decrease in the absorbance value, so the optimum temperature for study was 60°C ± 5°C (Figure 6). Effect of time The time of heating is an essential part of the experiment. Different time intervals were tested to ascertain the time after which the solution attains its highest absorbance. It was found that after 25 min, the reaction product reaches the highest absorbance (Figure 7). It was observed that heating time for 25 min is adequate and the absorbance of the reaction product is stable for about 40 min. at room temperature. The fluorescence intensity or absorbance value of the hydrolysis product of NBD-Cl, namely,4-hydroxy-7- nitrobenzo-2-oxa-1,3-diazole (NBD-OH) is quenched by decreasing the pH of the reaction medium to less than 1. Therefore, acidification of the reaction mixture prior to measurement of the fluorescence intensity or absorbance value remarkably decreased the background fluorescence or absorbance due to the formation of NBD- OH without affecting the drug reagent adduct, hence the sensitivity was increased. Absorbane 1.400 1.200 1.000 0.800 0.600 0.400 0.200 0.000 7.0 7.5 8.0 8.5 9.0 9.5 pH Figure 4 Effect of pH of borate buffer 0.2 M on the absorbance of the reaction product of SER (20 μg/mL) with NBD-Cl. Absorbane 1.400 1.200 1.000 0.800 0.600 0.400 0.200 0.000 0.0 0.5 1.0 1.5 2.0 Volume of 0.2% NBD-Cl (ml) Figure 5 Effect of volume of NBD-Cl 0.2% (w/v) on the absorbance of the reaction product of SER (20 μg/mL) with NBD-Cl.

Walash et al. Chemistry Central Journal 2011, 5:61 http://journal.chemistrycentral.com/content/5/1/61 Page 5 of 11 1.400 1.200 Analytical parameters for Sertraline Validation of the proposed methods The validity of the methods was tested regarding linearity, specificity, accuracy, repeatability and precision according to ICH Q2B recommendations [22]. Linearity The absorbance-concentration plot was linear over the range of 2-24 μg/mL with minimum detection limit (LOD) of 0.18 μg/mL. While, the fluorescenceconcentration plot was found to be linear over the range of 0.25-5 μg/mL with LOD of 0.07 μg/mL. Linear regression analysis of the data gave the following equations: Absorbane 1.000 0.800 0.600 0.400 A = 0.0389 + 0.0352 C (r = 0.9999) F = 19.038 + 31.814 C (r = 0.9999) 0.200 0.000 45.0 55.0 65.0 75.0 85.0 95.0 Temp. (°C) Figure 6 Effect of heating temperature on the absorbance of the reaction product of SER (20 μg/mL) with NBD-Cl. Effect of diluting solvent Dilution with different solvents such as methanol, water, acetone, dimethylsulfoxide and dimethylformamide was studied. The highest absorbance value was achieved upon diluting with methanol. Effect of surfactant The influence of concentration of different surfactants on the absorbance of the reaction product was studied hopefully they may enhance the absorbance. It was found that by using different concentrations of sodium dodecyl sulphate or cetrimide, there was a negligible increase in the absorbance, while using gelatine decreases the absorbance; therefore the study was performed omitting any surfactant. Absorbane 1.400 1.200 1.000 0.800 0.600 0.400 0.200 0.000 0.0 10.0 20.0 30.0 40.0 Time (min) Figure 7 Effect of heating time on the absorbance of the reaction product of SER (20 μg/mL) with NBD-Cl. Where A is the absorbance in 1-cm cell, F is the corrected fluorescence intensity, and C is the concentration of the drug in μg/mL. The limits of quantification (LOQ) and the limits of detection (LOD) were calculated according to ICH Q2B [22] using the following equations: LOQ = 10Sa/b LOD = 3.3Sa/b Where Sa = The standard deviation of the intercept of regression line b = Slope of the calibration curve. The results are shown in Table 1. LOQ were 0.56 and 0.21 μg/mL for methods I and II respectively, while LOD were 0.18 and 0.07 μg/mL for methods I and II respectively. These small values allowed the determination of SER in spiked human plasma using method II. The proposed methods were evaluated by studying the accuracy as percent relative error (% Er) (table 1) and precision as percent relative standard deviation (% RSD) and the results are shown in table 1. The small values of % Er and % RSD indicates high accuracy and high precision of the proposed methods. Accuracy To test the validity of the proposed methods, they were applied to the determination of pure sample of SER over the working concentration ranges. The results obtained were in good agreement with those obtained using reference method [23]. Using Student’s t-testand variance ratio F-test [25], revealed no significant difference between the performance of both methods and reference method regarding the accuracy and precision, respectively. The reference method recommended an HPLC determination of SER in pure and dosage forms using a mixture of 35% acetonitrile: methanol (92:8, v/v) and 65% of sodium acetate buffer at pH 4.5 as mobile phase with UV detection at 230 nm. The validity of the two methods were evaluated by statistical analysis of the regression lines regarding the standard deviation of the residuals (S y/x ), the standard deviation of the intercept (S a ) and standard deviation of the slope (S b ).TheresultsaregiveninTable1.The small values of the figures point out to the low scattering of the calibration graph and the precision of the methods.

Walash et al. Chemistry Central Journal 2011, 5:61<br />

http://journal.chemistrycentral.com/content/5/1/61<br />

Page 5 <strong>of</strong> 11<br />

1.400<br />

1.200<br />

Analytical parameters <strong>for</strong> Sertraline<br />

Validation <strong>of</strong> the proposed <strong>method</strong>s<br />

The validity <strong>of</strong> the <strong>method</strong>s was tested regarding linearity,<br />

specificity, accuracy, repeatability <strong>and</strong> precision according<br />

to ICH Q2B recommendations [22].<br />

Linearity<br />

The absorbance-concentration plot was linear over the<br />

range <strong>of</strong> 2-24 μg/mL with minimum detection limit<br />

(LOD) <strong>of</strong> 0.18 μg/mL. While, the fluorescenceconcentration<br />

plot was found to be linear over the range<br />

<strong>of</strong> 0.25-5 μg/mL with LOD <strong>of</strong> 0.07 μg/mL. Linear regression<br />

analysis <strong>of</strong> the data gave the following equations:<br />

Absorbane<br />

1.000<br />

0.800<br />

0.600<br />

0.400<br />

A = 0.0389 + 0.0352 C (r = 0.9999)<br />

F = 19.038 + 31.814 C (r = 0.9999)<br />

0.200<br />

0.000<br />

45.0 55.0 65.0 75.0 85.0 95.0<br />

Temp. (°C)<br />

Figure 6 Effect <strong>of</strong> heating temperature on the absorbance <strong>of</strong><br />

the reaction product <strong>of</strong> SER (20 μg/mL) with NBD-Cl.<br />

Effect <strong>of</strong> diluting solvent<br />

Dilution with different solvents such as methanol, water,<br />

acetone, dimethylsulfoxide <strong>and</strong> dimethyl<strong>for</strong>mamide was<br />

studied. The highest absorbance value was achieved<br />

upon diluting with methanol.<br />

Effect <strong>of</strong> surfactant<br />

The influence <strong>of</strong> concentration <strong>of</strong> different surfactants on<br />

the absorbance <strong>of</strong> the reaction product was studied hopefully<br />

they may enhance the absorbance. It was found that<br />

by using different concentrations <strong>of</strong> sodium dodecyl sulphate<br />

or cetrimide, there was a negligible increase in the<br />

absorbance, while using gelatine decreases the absorbance;<br />

there<strong>for</strong>e the study was per<strong>for</strong>med omitting any surfactant.<br />

Absorbane<br />

1.400<br />

1.200<br />

1.000<br />

0.800<br />

0.600<br />

0.400<br />

0.200<br />

0.000<br />

0.0 10.0 20.0 30.0 40.0<br />

Time (min)<br />

Figure 7 Effect <strong>of</strong> heating time on the absorbance <strong>of</strong> the<br />

reaction product <strong>of</strong> SER (20 μg/mL) with NBD-Cl.<br />

Where A is the absorbance in 1-cm cell, F is the corrected<br />

fluorescence intensity, <strong>and</strong> C is the concentration<br />

<strong>of</strong> the drug in μg/mL.<br />

The limits <strong>of</strong> quantification (LOQ) <strong>and</strong> the limits <strong>of</strong><br />

detection (LOD) were calculated according to ICH Q2B<br />

[22] using the following equations:<br />

LOQ = 10Sa/b LOD = 3.3Sa/b<br />

Where Sa = The st<strong>and</strong>ard deviation <strong>of</strong> the intercept <strong>of</strong><br />

regression line<br />

b = Slope <strong>of</strong> the calibration curve.<br />

The results are shown in Table 1.<br />

LOQ were 0.56 <strong>and</strong> 0.21 μg/mL <strong>for</strong> <strong>method</strong>s I <strong>and</strong> II<br />

respectively, while LOD were 0.18 <strong>and</strong> 0.07 μg/mL <strong>for</strong><br />

<strong>method</strong>s I <strong>and</strong> II respectively.<br />

These small values allowed the determination <strong>of</strong> SER<br />

in spiked human plasma using <strong>method</strong> II.<br />

The proposed <strong>method</strong>s were evaluated by studying the<br />

accuracy as percent relative error (% Er) (table 1) <strong>and</strong><br />

precision as percent relative st<strong>and</strong>ard deviation (% RSD)<br />

<strong>and</strong> the results are shown in table 1. The small values <strong>of</strong><br />

% Er <strong>and</strong> % RSD indicates high accuracy <strong>and</strong> high precision<br />

<strong>of</strong> the proposed <strong>method</strong>s.<br />

Accuracy<br />

To test the validity <strong>of</strong> the proposed <strong>method</strong>s, they were<br />

applied to the determination <strong>of</strong> pure sample <strong>of</strong> SER<br />

over the working concentration ranges. The results<br />

obtained were in good agreement with those obtained<br />

using reference <strong>method</strong> [23]. Using Student’s t-test<strong>and</strong><br />

variance ratio F-test [25], revealed no significant difference<br />

between the per<strong>for</strong>mance <strong>of</strong> both <strong>method</strong>s <strong>and</strong><br />

reference <strong>method</strong> regarding the accuracy <strong>and</strong> precision,<br />

respectively. The reference <strong>method</strong> recommended an<br />

HPLC determination <strong>of</strong> SER in pure <strong>and</strong> dosage <strong>for</strong>ms<br />

using a mixture <strong>of</strong> 35% acetonitrile: methanol (92:8, v/v)<br />

<strong>and</strong> 65% <strong>of</strong> sodium acetate buffer at pH 4.5 as mobile<br />

phase with UV detection at 230 nm.<br />

The validity <strong>of</strong> the two <strong>method</strong>s were evaluated by<br />

statistical analysis <strong>of</strong> the regression lines regarding the<br />

st<strong>and</strong>ard deviation <strong>of</strong> the residuals (S y/x ), the st<strong>and</strong>ard<br />

deviation <strong>of</strong> the intercept (S a ) <strong>and</strong> st<strong>and</strong>ard deviation <strong>of</strong><br />

the slope (S b ).TheresultsaregiveninTable1.The<br />

small values <strong>of</strong> the figures point out to the low scattering<br />

<strong>of</strong> the calibration graph <strong>and</strong> the precision <strong>of</strong> the<br />

<strong>method</strong>s.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!