Operating Manual (PDF) - John Morris Scientific
Operating Manual (PDF) - John Morris Scientific
Operating Manual (PDF) - John Morris Scientific
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Preparations<br />
3.2.1. External control<br />
Pt100<br />
The recirculating cooler is intended for internal and external temperature<br />
control. Switching is carried out in the submenu >Control< (see page 30)<br />
IMPORTANT: Measures for external control<br />
For external control and temperature measurement an external<br />
Pt100 sensor must be connected to the socket (17) on the rear of the<br />
recirculating cooler.<br />
Order No. Description Material cable<br />
8 981 003 200 x 6 mm dia., stainless steel 1.5 m<br />
8 981 005 200 x 6 mm dia., glass 1.5 m<br />
8 981 006 20 x 2 mm dia., stainless steel 1.5 m<br />
8 981 010 300 x 6 mm dia., stainless steel 1.5 m<br />
8 981 013 600 x 6 mm dia., stainless steel/Teflon 3 m<br />
8 981 014 1200 x 6 mm dia., stainless steel/Teflon 3 m<br />
8981103 Extension cable for Pt100 sensor 3.5 m<br />
8 981 020 M+R in-line Pt100 sensor<br />
The M+R in-line Pt100 sensor is a flow sensor and can be installed in<br />
the loop circuit.<br />
Pt100<br />
M+R<br />
3.3. Bath fluids<br />
Caution:<br />
Carefully read the safety data sheet of the bath fluid used, particularly with regard to<br />
the fire point!<br />
If a bath fluid with a fire point of ≤ 65 °C is used, only supervised operation is possible.<br />
Water:<br />
The quality of water depends on local conditions.<br />
Ferrous water can cause corrosion - even on stainless steel.<br />
Chloric water can cause pitting corrosion.<br />
Water/glycol mixture:<br />
Water evaporates and thus the mixture may become flammable. In the heating phase,<br />
a fire may be caused due to the hot surface temperature of the heater.<br />
Fill up water in time and check the mixing ratio in regular intervals (e.g. using an<br />
appropriate device).<br />
Recommended bath fluids:<br />
Bath fluids Temperature range Flash point fire point<br />
water 5 °C ... +80 °C<br />
mixture water/glycol -30 °C ... +50 °C<br />
JULABO Thermal G -30 °C ... +80 °C ---<br />
JULABO Thermal H5S -50 °C ... +105 °C >110 °C >130 °C<br />
18