12. Installation of electronic module<br />

ALARM STAND-BY<br />

Caution:<br />

Always turn off the unit and disconnect the mains cable from the power source before<br />

performing any service or maintenance procedures, or before moving the unit.<br />

Configuration, installation, maintenance and repairs on the circulator may only be<br />

carried out by qualified personnel.<br />

B Turn off the unit with the mains switch and disconnect the<br />

power plug.<br />

A<br />

REG+<br />

E-PROG<br />

<br />

<br />

<br />

<br />

Open the screws (A) and remove the dummy plate.<br />

Carefully insert the upper edge (B) into the rear panel from<br />

below, align the electronic module at top and bottom and<br />

applying light pressure connect the 15-pole connector.<br />

Attach the electronic module using the screws (A).<br />

The circulator is ready for use. It automatically recognizes the<br />

installed electronic module.<br />

13. Maintaining the cooling performance<br />

In dusty environments, the refrigerating machine’s condenser<br />

should be cleaned by vacuuming every 1 to 2 months. The<br />

condenser is located behind the venting grid on the front of the unit,<br />

which can be taken off by removing the screws.<br />

<br />

<br />

<br />

<br />

<br />

Switch off the unit, disconnect mains power cable.<br />

Remove the venting grid.<br />

Clean the ribbed condenser with a vacuum cleaner.<br />

Replace the venting grid.<br />

Switch on the unit.<br />