Troubleshooting guide / Error messages<br />
CONFIGURATION<br />
ERROR<br />
CONFIRM BY<br />
PRESSING : <br />
Special message „CONFIGURATION ERROR“<br />
The configuration of the circulator does not conform to its present use.<br />
Press enter to automatically perform a single modification of the<br />
configuration.<br />
Then contact an authorized JULABO service station.<br />
Disturbances that are not indicated.<br />
The electronic pump motor is overload-protected by an electronic current<br />
limiter. If viscosity of the bath fluid is or becomes too high, the motor<br />
stops running.<br />
Fuses:<br />
The mains fuses on the rear of the unit may easily be exchanged as<br />
shown on the left.<br />
Circulator: Fine fuses (15)– T 1.25 A, 250 V~ , D 5 x 20 mm<br />
Circulator: Fine fuses (16a)– T 16 A, 250 V~ , D5 x 20 mm<br />
Warning:<br />
Before exchanging the fuses, turn off the mains power switch and disconnect the<br />
power plug from the mains socket!<br />
Only use fine fuses with a nominal value as specified.<br />
Example:<br />
Manufacturer Supplier Type Order No.<br />
Schurter Schurter G-fuse insert SPT<br />
No. 0001.2516<br />
T16A 5x20mm<br />
Wickmann Wickmann G-fuse insert<br />
T1.25A 5x20 mm<br />
No. 19195<br />