Troubleshooting guide / Error messages<br />

ALARM<br />

CODE 05<br />

Internal working<br />

sensor shurtcircuited<br />

or<br />

interrupted.<br />

<br />

Cable of the working temperature sensor interrupted or shortcircuited.<br />

ALARM<br />

CODE 06<br />

Difference between<br />

working- and safetytemp.<br />

Bigger than<br />

permissable value.<br />

<br />

Defect of the working or excess temperature sensor.<br />

Working temperature and excess temperature sensors report a<br />

temperature difference of more than 25 °C.<br />

ALARM<br />

CODE 07<br />

The internal I2C<br />

bus-system shows<br />

a failure.<br />

<br />

Other errors (I 2 C-BUS errors)<br />

ALARM<br />

CODE 12<br />

Error in A/Dconverter<br />

of the<br />

internal measuring<br />

system.<br />

<br />

Error in A/D converter<br />

ALARM<br />

CODE 14<br />

Working temperature<br />

exceeds the adjusted<br />

safety temperature.<br />

<br />

<br />

Excess temperature sensors defect.<br />

The excess temperature value lies below the working temperature<br />

setpoint. Set the excess temperature to a higher value.<br />

ALARM<br />

CODE 15<br />

External control<br />

selected, but ext.<br />

sensor not connected<br />

or defective.<br />

<br />

External control selected, but external Pt100 sensor not connected or<br />

defect.<br />

WARNING<br />

CODE 20<br />

<br />

Cooling of the condenser is affected. Clean air-cooled condenser.<br />

Check the flow rate and cooling water temperature on water-cooled<br />

condenser.<br />