Menu functions<br />
6.9. Setting the pump pressure<br />
Examples:<br />
Adjusted<br />
Effective<br />
The pressure of the circulating pump is adjustable in four grades. After<br />
setting, the VFD-Info-Display indicates the corresponding value.<br />
Adjustable pump capacity stage 1 ... 4<br />
Illuminated display:<br />
for pump pressure<br />
Display for the adjusted pump pressure stage in the –OFF- mode.<br />
Display for the effective pump pressure stage (rotation speed) after start.<br />
For protecting the pump motor, the rotation speed changes with the load<br />
applied, for example depending on the viscosity of the bath fluid at different<br />
working temperatures.<br />
Flow rate:<br />
22 ... 26 l/min<br />
Pump capacity stage 1 2 3 4<br />
Pump pressure: [bar] 0,4 0,5 0,6 0,7<br />
Suction pump [bar] 0,2 0,26 0,33 0,4<br />
Total capacity [bar] 0,6 0,76 0,93 1,1<br />