Menu functions<br />

Type:<br />

A >1-point2-point < or >3-point < calibration can be carried<br />

out.<br />

First geometrically define the location for calibration (measuring point<br />

T M ), then determine the temperature values of of the calibration<br />

points.<br />

The calibration type also determines the number of pairs of values<br />

indicated on the LCD DIALOG-DISPLAY.<br />

TmpValue: 1 or 2 or 3<br />

Defined temperature value of the calibration point. This value is<br />

automatically stored with >CalValue< and can be indicated for control<br />

purposes.<br />

CalValue: 1 or 2 or 3<br />

The „Calibration value“ is determined with a temperature measuring<br />

device and stored under menu item >CalValueATCalibration< under menu item >Calwert X