Ultra-Low Refrigerated Circulators<br />

6.8. ATC Absolute Temperature Calibration, 3-point calibration<br />

ATC serves to compensate a temperature difference that might occur between<br />

circulator and a defined measuring point in the bath tank because of physical<br />

properties.<br />

Limits<br />

Interface serial<br />

>ATCalibration<br />

Pump<br />

ESC<br />

ESC<br />

ATCalibration<br />

>Sensor : intern<br />

Status : no /yes<br />

Type : 3-point /1-point/2-point<br />

TmpValue1: xx.xx°C<br />

CalValue1: xx.xx°C<br />

TmpValue2: xx.xx°C<br />

CalValue2: xx.xx°C<br />

TmpValue3: xx.xx°C<br />

1<br />

CalValue3: xx.xx°C<br />

internal<br />

ESC<br />

ATCalibration<br />

>Sensor : extern<br />

Status : no /yes<br />

Type : 3-point /1-point/2-point<br />

TmpValue1: xx.xx°C<br />

CalValue1: xx.xx°C<br />

TmpValue2: xx.xx°C<br />

CalValue2: xx.xx°C<br />

external<br />

1<br />

TmpValue3: xx.xx°C<br />

CalValue3: xx.xx°C<br />

Sensor: >internextern<<br />

Calibration can be carried out for the internal temperature sensor and<br />

for the external temperature sensor connected to the socket „ext.<br />

Pt100“.<br />

The circulator is able to store both parameter sets. However, only the<br />

one set under this menu item is indicated.<br />

Status:<br />

>no< The controller of the circulator uses the original curve of<br />

the temperature sensor.<br />

Important: During the calibration process >no< needs to be set.<br />

>yes<<br />

curve.<br />

The controller of the circulator uses the new calibration<br />

59