Ultra-Low Refrigerated Circulators<br />

>SafeTemp:<<br />

IntMax: and IntMin: Limit setting<br />

Setting range: see working temperature range<br />

The limits IntMax and IntMin are only valid for external control. IntMax and<br />

IntMin are used to limit the expected internal bath temperatures to any upper<br />

and lower values. The temperature controller cannot exceed these limits even if<br />

it would be required for reaching the temperature in the external system.<br />

Consequently the external setpoint may thus not be reached.<br />

Sense of a limit setting:<br />

Protects the bath fluid from overheating.<br />

Prevents an undesired alarm shut-down by the excess temperature<br />

protection - >ALARM CODE 14IntMax:< to a value at least 5 °C below the >SafeTemp:< value.<br />

Protects the pump motor from high viscosity of the bath fluid at low<br />

temperatures.<br />

For refrigerated circulators: Freezing protection when using water as bath<br />

fluid.<br />

°C<br />

200<br />

150<br />

Upper band: and Lower band: Band limit<br />

The band limit is active for external control.<br />

internal bath<br />

external system<br />

For the heat-up and cool-down phases<br />

different settings to conform to the<br />

requirements of the particular application<br />

are possible.<br />

Setting range: 0 ... 200 °C<br />

100<br />

75<br />

50<br />

25<br />

AreaUp<br />

AreaDown<br />

10 20 30 40 50 60 10 20 30 40 50 60 10 [min]<br />

During the heat-up phase the difference value always adds to the actual<br />

external temperature. During the cool-down phase, the difference value is<br />

subtracted.<br />

t<br />

Using Upper band and Lower band, the<br />

difference between the temperatures in<br />

the internal bath and the external system<br />

can be limited to any maximum value for<br />

the heat-up or the cool-down phase.<br />

Sense of a band limit:<br />

Ensures gentle heating for the objects requiring temperature application.<br />

Protects glass reactors, for example, from thermal shock.<br />

The IntMax: and IntMin: values are superordinated to this band limit.<br />

57