Menu functions<br />

6.6. Limits<br />

The menu >Limits< allows for setting the minimum and maximum values for all<br />

important setting ranges and capacity variables.<br />

Inputs/Outputs<br />

>Limits<br />

Interface serial<br />

ATCalibration<br />

ESC<br />

ESC<br />

Limits<br />

>SetMax :300.00°C<br />

SetMin :-94.99°C<br />

HeatingMax: 100%<br />

CoolingMax: 0%<br />

IntMax :300.00°C<br />

IntMin :-94.99°C<br />

Upper band: 200K<br />

Lower band: 200K<br />

1<br />

In case of external control<br />

these menu items are<br />

additionally indicated<br />

SetMax: and SetMin:<br />

Maximum and minimum setpoint<br />

Restriction for the adjustable temperature range.<br />

Setting range: see working temperature range<br />

The limitation of the operating temperature range effects the temperature<br />

setting under the menu called up with the key - Setpoint 1, 2, 3.<br />

It is possible to adjust only working temperatures that lie within the limit range<br />

set here.<br />

Existing settings for Setpoint 1, 2, 3 and also for >OverTemp< and >SubTemp<<br />

(see page 28) are automatically defered within the limit range.<br />

HeatingMax:<br />

and CoolingMax:<br />

Adjusted maximum heating / cooling.<br />

Setting range: 0 to 100 % in steps of 1 %<br />

Heating and cooling powers of the circulator are adjustable. 100 %<br />

corresponds to the values in the technical specifications of the equipment.<br />

Factory setting: CoolingMax<br />

Refrigorated/heating circulator Fxx = 0 % eg. F81<br />

Refrigorated/heating circulator FPxx = 100 % eg. FP88<br />

56