Menu functions<br />

Menu >Configuration<<br />

Configuration<br />

>Setpoint :<br />

Autostart: off<br />

Standby : pump off<br />

Language : english<br />

Reset : no<br />

Variable : Control /serial/eprog<br />

eprog Keypad/RS232/extPt100<br />

E-PROG - Input<br />

Setting needs to be carried out, if<br />

1. setpoint programming is to be made via an<br />

external programmer.<br />

For this, in the menu >Configuration< first set<br />

the menu item >Setpoint< to >eprogInputs/Outputs<<br />

>EPROG : Flowrate<br />

Signal : Voltage<br />

L Value 0.0L<br />

H Value 50.0L<br />

Flowrate: xx.xL<br />

Setpoint/Variable<br />

(Liters/Minute)<br />

2. the heater variable should be controlled via an<br />

external control pulse.<br />

For this, in the menu >Configuration< set the<br />

menu item > Variable.< to >eprogSetpoint< or under menu item > Variable Setpoint< or > Variable ConfigurationInputs/OutputsEPROG : Setpoint<br />

Signal : Current<br />

L Value 0.0°C<br />

H Value 300.0°C<br />

Setpoint : 50.0°C<br />

<br />

Connect the external programmer to socket (12) REG+E-PROG of<br />

the circulator.<br />

Selecting the signal:<br />

The programmer (E-PROG) input of the circulator can be matched to<br />

the output signal of the external programmer.<br />

Voltage voltage input<br />

Current current input<br />

„L Value“ - Setting the LOW value: (See below )<br />

Adjust and set the lowest desired working temperature on the<br />

programmer (e.g. 0 C).<br />

Enter the same value on the circulator by pressing the appropriate<br />

buttons on the keypad and press enter<br />

to set.<br />

„H Value“ - Setting the HIGH value: (See below )<br />

Adjust and set the highest desired working temperature on the<br />

programmer (e.g. 300 C).<br />

Enter the same value on the circulator by pressing the appropriate<br />

buttons on the keypad and press enter<br />

to set.<br />

Example:<br />

Set a temperature of 50 C on the external programmer!<br />

The value adjusted and set on the external programmer is displayed<br />

in line 4 of the LCD DIALOG-DISPLAY for control purposes.<br />

52