Ultra-Low Refrigerated Circulators<br />

REG+E-PROG (12):<br />

Outputs of the connector<br />

1. First define the desired output value for channels 1 to 3:<br />

ActInt internal actual temperature value (bath temperature)<br />

ActExt ernal actual temperature value (external sensor)<br />

Power periodic or intermittent heating or cooling<br />

Setpoint active setpoint temperature<br />

(Setpoint1, 2, 3,/ integr. programmer /external programmer)<br />

2. Then select the display size for channels 1 to 3:<br />

Voltage outputs channels 1 and 2<br />

Assign the voltage values of 0 V to the lowest and 10 V to the highest<br />

necessary temperature or power rating required as an output value<br />

(°C / %).<br />

Current output channel 3<br />

Assign the current values 0 mA or 4 mA to the lowest and 20 mA to<br />

the highest temperature or power rating required as an output value<br />

(°C / %).<br />

3. The current output (chanell 3) offers 2 ranges for selection:<br />

0 to 20 mA or 4 to 20 mA<br />

0 V<br />

10 °C 210 °C<br />

200 K<br />

197 °C 202 °C<br />

5 K<br />

0 mA 20 mA<br />

4 mA<br />

Examples:<br />

>Chan.1 : ActInt<br />

0V = 10°C<br />

10V = 210°C<br />

lowest temperature value: 10 °C<br />

highest temperature value 210 °C<br />

Fig. shows 200 C scaled to paper width<br />

rise: 50 mV/°C<br />

lowest temperature value: 197 °C<br />

highest temperature value: 202 °C<br />

Fig. shows 5 C scaled to paper width<br />

rise: 2000 mV/°C<br />