Ultra-Low Refrigerated Circulators<br />
REG+E-PROG (12):<br />
Outputs of the connector<br />
1. First define the desired output value for channels 1 to 3:<br />
ActInt internal actual temperature value (bath temperature)<br />
ActExt ernal actual temperature value (external sensor)<br />
Power periodic or intermittent heating or cooling<br />
Setpoint active setpoint temperature<br />
(Setpoint1, 2, 3,/ integr. programmer /external programmer)<br />
2. Then select the display size for channels 1 to 3:<br />
Voltage outputs channels 1 and 2<br />
Assign the voltage values of 0 V to the lowest and 10 V to the highest<br />
necessary temperature or power rating required as an output value<br />
(°C / %).<br />
Current output channel 3<br />
Assign the current values 0 mA or 4 mA to the lowest and 20 mA to<br />
the highest temperature or power rating required as an output value<br />
(°C / %).<br />
3. The current output (chanell 3) offers 2 ranges for selection:<br />
0 to 20 mA or 4 to 20 mA<br />
0 V<br />
10 °C 210 °C<br />
200 K<br />
197 °C 202 °C<br />
5 K<br />
0 mA 20 mA<br />
4 mA<br />
Examples:<br />
>Chan.1 : ActInt<br />
0V = 10°C<br />
10V = 210°C<br />
lowest temperature value: 10 °C<br />
highest temperature value 210 °C<br />
Fig. shows 200 C scaled to paper width<br />
rise: 50 mV/°C<br />
lowest temperature value: 197 °C<br />
highest temperature value: 202 °C<br />
Fig. shows 5 C scaled to paper width<br />
rise: 2000 mV/°C<br />
51