Ultra-Low Refrigerated Circulators<br />

6.5. Analog inputs/outputs<br />

ALARM STAND-BY<br />

REG+<br />

E-PROG<br />

In order to use the analog inputs and outputs, the circulator must be<br />

equipped with the electronics module available as option.<br />

Order No. 8 900 100 Electronic module<br />

This submenu enables setting of the input and output values for the<br />

programmer input and the temperature recorder outputs of socket<br />

REG+E-PROG.<br />

The >STAND-BY< input and the >ALARM< output are configurable.<br />

If the electronic module is not fitted, this menu can only be used to<br />

switch control of the solenoid valve.<br />

Int. Programmer<br />

>Input/Output<br />

Limits<br />

Interface serial<br />

ESC<br />

Indication without electronic module<br />

Inputs/Outputs<br />

>SV-Out. : Backflow /C-Puls<br />

ESC<br />

Inputs/Outputs<br />

>SV-Out. : Backflow /C-Puls<br />

Chan.1 : ActInt<br />

Chan.2 : Power<br />

Chan.3 : Setpoint<br />

EPROG : Setpoint<br />

Ext.StBy : inactive /active<br />

AlarmOut : StdBy<br />

>Chan.1 : ActInt /<br />

0V = -99.9°C<br />

10V = 400.0°C<br />

ActExt/Power/Setpoint<br />

1<br />

>Chan.2 : Power /<br />

0V = 0.00%<br />

10V = 100%<br />

ActInt/ActExt/Setpoint<br />

1<br />

ESC<br />

>Chan.3 : Setpoint / ActExt/ACtInt/Power<br />

0mA = -99.9°C<br />

20mA = 400.0°C 1<br />

Region: 0-20mA /4-20mA<br />

>EPROG : Setpoint / ActVar (%)/Flowrate(l/min)<br />

Signal : Voltage /Current<br />

L Value -99.9°C REM: % / l/Min<br />

H Value 400.0°C REM: % / l/Min<br />

Setpoint : xx.xx°C REM: % / l/Min 1<br />

AlarmOut<br />

>Function: StdBy /Alarm/Al+StBy<br />

Type : normal /invers<br />

NOTE: Channel 2 is pre-configured for connecting a HST<br />

Booster heater.<br />

49