Menu functions<br />

6.3.2. Interrupting a profile<br />

Setp. : 125.80°C<br />

IntAct: 124.70°C<br />

Prof. : 4 Step:12<br />

*** pause ***<br />

Press the start/stop key to interrupt or continue a profile.<br />

The setpoint and time period set for the corresponding section are<br />

thus stopped at the values presently achieved.<br />

The instrument is put on hold and the message „pause“ flashes on<br />

the LCD DIALOG DISPLAY.<br />

STAND-BY<br />

Setp. : xxx.xx°C<br />

IntAct: xxx.xx°C<br />

Prof. : 4 Step:12<br />

*** STAND-BY ***<br />

A profile can be interrupted or restarted by an external emergency<br />

shut-off.<br />

Important:<br />

For this, in the menu >Inputs/Outputs< the menu item >Ext. StBy <<br />

must be set to >activ< (see page 54)<br />

and in the menu >Configuration< >Autostart< needs to be set (see<br />

page 33).<br />

Setpoint control and timer are interrupted by breaking the contact<br />

„AK”.<br />

The instrument is put on hold and the message „Stand-By“ flashes on<br />

the LCD DIALOG-DISPLAY.<br />

Notice: This is not an actual emergency switch-off.<br />

Warning:<br />

Following a power interruption, it would be possible in this condition for the instrument<br />

to restart automatically. The safety and warning functions of the instrument should<br />

always be used to their fullest capacity.<br />

See Warning page 33.<br />

6.3.3. Interruption after a power failure<br />

There is every possibility to operate the programmer safe to power<br />

failures.<br />

Important:<br />

For this in the menu >Configuration< >Autostart< needs to be set (see<br />

page 33).<br />

Starttime: 06:00<br />

Date : 15:12<br />

*** wait ***<br />

Acttime 02:34:45<br />

If the AUTOSTART function is activated, the programmer starts again<br />

at a point approx. 20 seconds before the interruption took place.<br />

However, an uncontrolled change of the bath temperature happened.<br />

Power failure when the instrument is on hold if start is effected via the<br />

built-in timer:<br />

If the starting time is not yet reached, the programmer is put on hold<br />

again.<br />

If the starting time is exceeded, the programmer starts immediately.<br />