Menu functions<br />

Dynamic:<br />

°C<br />

normal<br />

temp. stability<br />

This parameter affects the temperature pattern only in case<br />

of internal control.<br />

aperiodic<br />

Possible parameters:<br />

setpoint<br />

normal –Allows for reaching the setpoint faster, but<br />

overshooting of up to 5 % is possible.<br />

°C<br />

normal<br />

t<br />

Aperiodic – (factory setting)<br />

Target temperature is attained without overshooting.<br />

temp. stability<br />

aperiodic<br />

With both adjustments an adequate temperature<br />

stability is reached after approximately the same time.<br />

temperature ramp<br />

t<br />

CoSpeed:<br />

This parameter affects the temperature pattern only in case<br />

of external control.<br />

Possible parameters: 0 ... 5<br />

°C<br />

Int. temperature<br />

Ext. temperature<br />

5<br />

3<br />

0<br />

Setpoint<br />

t<br />

During selftuning, the control parameters Xp, Tn<br />

and Tv of a controlled system are automatically<br />

determined and stored. Depending on the<br />

controlled system, time for tuning can be<br />

unequally longer. This controller layout allows<br />

protection of sensitive objects requiring<br />

temperature application.<br />

As soon as a co-speed factor is set, it is<br />

considered for calculating the control parameters.<br />

As shown in the diagram, tuning times become<br />

shorter the higher the co-speed factor is, but<br />

overshooting can happen in the internal system.<br />

Control parameters – intern / extern<br />

The control parameters preset in factory are in most cases adequate for<br />

achieving an optimum temperature pattern for the samples requiring<br />

temperature application.<br />

Each parameter may be manually set via the keypad if necessary, to allow<br />

optimum control performance.<br />

Setting range:<br />

internal / external<br />

0.1 ... 99.9 K<br />

Proportional range >Xp<<br />

The proportional range is the range below the selected temperature value in<br />

which the control circuit reduces the heating power from 100 % to 0 %.<br />

38