Ultra-Low Refrigerated Circulators<br />

-OFF-<br />

S xx.xx<br />

E xx.xx<br />

ControlType: Can only be adjusted in the -OFF- condition<br />

Possible parameters:<br />

intern internal temperature control<br />

extern external temperature control with external Pt100 sensor<br />

ext.<br />

Pt100<br />

IMPORTANT:<br />

Additional measures for external temperature control.<br />

Connect a Pt100 sensor to the socket on the rear of the circulator.<br />

Sensor calibration of the external Pt100 sensor is carried out in the menu<br />

>ATCalibration Status< set to >no< (see page 59).<br />

Suggested adjustments for external temperature control:<br />

AreaUp/ AreaDown and IntMax / IntMin<br />

see chapter >Limits< on page 56.<br />

External temperature control does not allow for setpoint setting via the<br />

socket „ext. Pt100“ (see page 32).<br />

Notice:<br />

Place the external sensor into the bath medium and securely fix the sensor.<br />

Selftuning:<br />

When performing an selftuning for the controlled system (temperature<br />

application system), the control parameters Xp, Tn and Tv are automatically<br />

determined and stored.<br />

Possible parameters:<br />

off - no selftuning.<br />

The control parameters ascertained during the last identification are used<br />

for control purposes.<br />

once – single selftuning (factory setting)<br />

The instrument performs a single selftuning of the controlled system after<br />

each start with the start/stop key or after receiving a start command<br />

via the interface.<br />

always – continual selftuning<br />

The instrument performs a selftuning of the controlled system whenever<br />

a new setpoint is to be reached.<br />

Use this setting only when the temperature application system changes<br />

permanently.<br />

37