Ultra-Low Refrigerated Circulators<br />

Warning:<br />

The working temperature range of the refrigerated circulator is determined during<br />

configuration. If set to >Control seriel < and > eprog serial< and >eprog<<br />

Start conditions:<br />

The external variable lies within the range –1 % to –100 %.<br />

<br />

<br />

Time limit:<br />

A 1-stage cooling machine produces cooling capacity after 5 minutes, a 2-stage<br />

cooling machine after 10 minutes.<br />

The indicator light illuminates only after this delay.<br />

This time delay after switch-off and switch-on of the cooling unit ensures protection<br />

of the cooling compressor.<br />

Temperature limit<br />

The bath temperature must be within the configured range (working temperature<br />

range) of the refrigerated circulator. If the temperature exceeds the maximum<br />

temperature, the compressor remains switched-off if cooling is requested. If the<br />

bath temperature falls below the configured temperature range, the compressor<br />

switches on and cold is applied to the bath after the time delay indicated above.<br />

Stop conditions:<br />

<br />

<br />

If the external variable is above 95 % (heating capacity) for more than 5 minutes,<br />

the cooling machine is switched off.<br />

Refrigerated circulators are configured to the effect that the compressor can only<br />

operate within a certain, limited temperature range. The compressor is switched off<br />

if the temperature exceeds the configured maximum temperature.<br />

Special conditions on FP88-HL<br />

If cooling is requested, the FP88 cooling machine changes the variable to the effective<br />

value within 100 seconds starting at 0 % (at –100 % this means –1 % per second).<br />

Only after this time is elapsed, cooling capacity is produced and the indicator light<br />

illuminates.<br />

If the variable is changed to the heating range (between 1 % to 100 %) within these<br />

100 seconds, then the time is set to 0 if cooling is requested again.<br />

35