Operating Manual - John Morris Scientific
Operating Manual - John Morris Scientific
Operating Manual - John Morris Scientific
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Ultra-Low Refrigerated Circulators<br />
eprog : 50.0°C<br />
IntAct : 24.7°C<br />
Power : 80 %<br />
Control : intern<br />
eprog – Can only be adjusted when an electronic module with analog<br />
connections is used (option).<br />
Setpoint setting via the analog interface REG+E-PROG through an external<br />
programmer.<br />
Important:<br />
Connect the external programmer to the circulator via the socket REG+E-<br />
PROG.<br />
In the menu >Inputs/Outputs< set the parameter >EPROG< and the input<br />
variables (see page 52).<br />
The E-Prog input can only be used either under menu item >Setpoint< or<br />
under menu item > Variable < (page 34).<br />
Autostart<br />
Possible parameters:<br />
on - AUTOSTART on<br />
off - AUTOSTART off (Factory setting)<br />
Note:<br />
The temperature system has been configured and supplied by JULABO<br />
according to N.A.M.U.R. recommendations. This means for the start mode, that<br />
the unit must enter a safe operating state after a power failure (non-automatic<br />
start mode). This safe operating state is indicated by „ OFF „, resp. R OFF on<br />
the VFD-COMFORT-DISPLAY. A complete shutdown of the main functional<br />
elements such as heater and circulating pump is effected simultaneously.<br />
The values set on the circulator remain stored, and the unit is returned to<br />
operation by pressing the start/stop key (in manual control mode).<br />
In remote control mode, the values need to be resent by the PC via the<br />
interface.<br />
Should such a safety standard not be required, the AUTOSTART function<br />
(automatic start mode) may be activated, thus allowing the start of the<br />
instrument directly by pressing the mains power switch or using a timer.<br />
The AUTOSTART function can only be used, if set point setting is carried<br />
out via >keyboard< or >eprog< or >ext. Pt100