Menu functions<br />
6.1. Configuration<br />
>Configuration<br />
Control<br />
Profile Start<br />
Int.Programmer<br />
ESC<br />
ESC<br />
Configuration<br />
>Setpoint : keypad /RS232/ext.Pt100/eprog<br />
Autostart: off /on<br />
OFF-Mode : pump off /pump on<br />
Language : German /Englisch<br />
Reset : no /yes<br />
Variable : control /serial/eprog<br />
Unit : °C /°F<br />
1. Use the up/down keys to select the desired menu item.<br />
2. Use the left/right keys to select the parameter (parameter blinks).<br />
3. Press enter to store the new parameter.<br />
Setpoint<br />
The circulator provides four possibilities for setpoint setting. The selected mode<br />
is indicated in line 1 on the LCD DIALOG-DISPLAY.<br />
Setpoint1: 150.00°C<br />
IntAct : 148.53°C<br />
Power : 80 %<br />
Control : intern<br />
RS232 : 150.00°C<br />
ExtAct : 145.30°C<br />
Power : 80 %<br />
Control : extern<br />
R<br />
-OFF-<br />
S xxxx<br />
I xxxx<br />
Possible parameters:<br />
Keypad – Setpoint setting via keypad or via the integrated programmer.<br />
(factory setting)<br />
RS232 or RS485 - Setpoint setting via the serial RS232/RS485 interface (20)<br />
through a PC or superordinated data system.<br />
In the header of the VFD COMFORT-DISPLAY an „R“ illuminates. It<br />
indicates that remote control mode is set.<br />
Important:<br />
Connect the circulator to a PC using an interface cable.<br />
Check the interface parameters of both interfaces (on circulator and PC)<br />
and make sure they match.<br />
(see 10.1. Setup for remote control on page 71 )<br />
Pt100 : 150.00°C<br />
ExtAct : 145.30°C<br />
Power : 80 %<br />
Control : extern<br />
ext.Pt100 – Setpoint setting via the analog socket „ext. Pt100“ using an<br />
external temperature sensor or an appropriate voltage/current source.<br />
32