Menu functions<br />

6.1. Configuration<br />

>Configuration<br />

Control<br />

Profile Start<br />

Int.Programmer<br />

ESC<br />

ESC<br />

Configuration<br />

>Setpoint : keypad /RS232/ext.Pt100/eprog<br />

Autostart: off /on<br />

OFF-Mode : pump off /pump on<br />

Language : German /Englisch<br />

Reset : no /yes<br />

Variable : control /serial/eprog<br />

Unit : °C /°F<br />

1. Use the up/down keys to select the desired menu item.<br />

2. Use the left/right keys to select the parameter (parameter blinks).<br />

3. Press enter to store the new parameter.<br />

Setpoint<br />

The circulator provides four possibilities for setpoint setting. The selected mode<br />

is indicated in line 1 on the LCD DIALOG-DISPLAY.<br />

Setpoint1: 150.00°C<br />

IntAct : 148.53°C<br />

Power : 80 %<br />

Control : intern<br />

RS232 : 150.00°C<br />

ExtAct : 145.30°C<br />

Power : 80 %<br />

Control : extern<br />

R<br />

-OFF-<br />

S xxxx<br />

I xxxx<br />

Possible parameters:<br />

Keypad – Setpoint setting via keypad or via the integrated programmer.<br />

(factory setting)<br />

RS232 or RS485 - Setpoint setting via the serial RS232/RS485 interface (20)<br />

through a PC or superordinated data system.<br />

In the header of the VFD COMFORT-DISPLAY an „R“ illuminates. It<br />

indicates that remote control mode is set.<br />

Important:<br />

Connect the circulator to a PC using an interface cable.<br />

Check the interface parameters of both interfaces (on circulator and PC)<br />

and make sure they match.<br />

(see 10.1. Setup for remote control on page 71 )<br />

Pt100 : 150.00°C<br />

ExtAct : 145.30°C<br />

Power : 80 %<br />

Control : extern<br />

ext.Pt100 – Setpoint setting via the analog socket „ext. Pt100“ using an<br />

external temperature sensor or an appropriate voltage/current source.<br />