Ultra-Low Refrigerated Circulators<br />

Setpoint3 85.0°C<br />

OverTemp 87.0°C<br />

SubTemp 83.0°C<br />

The warning functions are only triggered when the actual bath temperature,<br />

after start from the „OFF“ or „rOFF“ mode, lies within the set limits for 3<br />

seconds.<br />

Setting range:<br />

20 °C ... 320 °C<br />

SafeTemperature: Indicated is the cut-out value set with a screwdriver on the<br />

excess temperature protection device.<br />

This safety installation is independent of the control circuit. When the<br />

temperature of the bath fluid has reached the safety temperature, a complete<br />

shutdown of the heater and pump is effected.<br />

The alarm is indicated by optical and audible signals (continuous tone).<br />

The following error message appears on the VFD COMFORT-DISPLAY:<br />

XXXXX<br />

ALARM<br />

CODE 14<br />

A L A R M !<br />

IntAct: 24.60°C<br />

Safety circuit alarm<br />

Press<br />

for help on the LCD<br />

Working temperature<br />

exceeds the adjusted<br />

safety temperature.<br />

<br />

SafeSens: Indicated is the temperature value of the safety sensor.<br />