<strong>Manual</strong> operation<br />

Recommendation:<br />

Set the excess temperature protection >SafeTemp< at 5 to 10 °C above the<br />

working temperature setpoint.<br />

Warning:<br />

The excess temperature protection >SafeTemp< should be set at least 25 °C below<br />

the fire point of the bath fluid used.<br />

In the event of wrong setting there is a fire hazard!<br />

We disclaim all liability for damage caused by wrong settings!<br />

Warn-Type: >Warning< or >Alarm<<br />

For the two menu items >OverTemp< and >SubTemp< choose between a<br />

warning message being signaled or a complete shutdown of the main<br />

functional elements such as heater and circulating pump being effected.<br />

OverTemperature: SubTemperature:<br />

If for a sensitive temperature application task adherence to a working<br />

temperature value >Setpoint< is to be supervised, then set high and low<br />

temperature warning values.<br />

In the example below, the >Setpoint< of 85 °C is surrounded by the values<br />

>OverTemp< 87 °C and >SubTemp< 83 °C. The electronics immediately<br />

registers when the actual temperature attains a temperature out of the<br />

limits and it follows a reaction according to what is set in the menu item<br />

>Warn-TypeWarning<<br />

An audible warning (interval tone)<br />

sounds and a meassage appears on<br />

the VFD COMFORT-DISPLAY.<br />

OverTemp SubTemp<br />

XXXXX<br />

WARNING<br />

CODE 03<br />

XXXXX<br />

WARNING<br />

CODE 04<br />

Warn-Type: >Alarm<<br />

A complete shutdown of heater and<br />

circulating pump is effected.<br />

An audible alarm (continuous tone)<br />

sounds and a message appears on the<br />

VFD COMFORT-DISPLAY.<br />

OverTemp SubTemp<br />

XXXXX<br />

ALARM<br />

CODE 03<br />

XXXXX<br />

ALARM<br />

CODE 04<br />

Press for help on the LCD.<br />

Actual temperature<br />

above high temperature<br />

value.<br />

Actual temperature<br />

below low temperature<br />

value.<br />