<strong>Manual</strong> operation<br />
Recommendation:<br />
Set the excess temperature protection >SafeTemp< at 5 to 10 °C above the<br />
working temperature setpoint.<br />
Warning:<br />
The excess temperature protection >SafeTemp< should be set at least 25 °C below<br />
the fire point of the bath fluid used.<br />
In the event of wrong setting there is a fire hazard!<br />
We disclaim all liability for damage caused by wrong settings!<br />
Warn-Type: >Warning< or >Alarm<<br />
For the two menu items >OverTemp< and >SubTemp< choose between a<br />
warning message being signaled or a complete shutdown of the main<br />
functional elements such as heater and circulating pump being effected.<br />
OverTemperature: SubTemperature:<br />
If for a sensitive temperature application task adherence to a working<br />
temperature value >Setpoint< is to be supervised, then set high and low<br />
temperature warning values.<br />
In the example below, the >Setpoint< of 85 °C is surrounded by the values<br />
>OverTemp< 87 °C and >SubTemp< 83 °C. The electronics immediately<br />
registers when the actual temperature attains a temperature out of the<br />
limits and it follows a reaction according to what is set in the menu item<br />
>Warn-TypeWarning<<br />
An audible warning (interval tone)<br />
sounds and a meassage appears on<br />
the VFD COMFORT-DISPLAY.<br />
OverTemp SubTemp<br />
XXXXX<br />
WARNING<br />
CODE 03<br />
XXXXX<br />
WARNING<br />
CODE 04<br />
Warn-Type: >Alarm<<br />
A complete shutdown of heater and<br />
circulating pump is effected.<br />
An audible alarm (continuous tone)<br />
sounds and a message appears on the<br />
VFD COMFORT-DISPLAY.<br />
OverTemp SubTemp<br />
XXXXX<br />
ALARM<br />
CODE 03<br />
XXXXX<br />
ALARM<br />
CODE 04<br />
Press for help on the LCD.<br />
Actual temperature<br />
above high temperature<br />
value.<br />
Actual temperature<br />
below low temperature<br />
value.<br />
28