Ultra-Low Refrigerated Circulators<br />

5.3. Safety installations, warning functions<br />

Settings for the excess temperature protection according to IEC 61010-2-010<br />

and for the high and low temperature warning functions are made<br />

in a menu that is called up with the key .<br />

ESC<br />

Safety-Values<br />

>Warn-Type: Warning /Alarm<br />

OverTemp: 85.00°C<br />

SubTemp : -99.99°C<br />

SafeTemp : 100.00°C<br />

SafeSens : xx.xx°C<br />

1<br />

80<br />

40<br />

0<br />

160<br />

240<br />

320<br />

°C°C<br />

Legend:<br />

1<br />

Setting via keypad<br />

Select parameter<br />

Select menu item<br />

Change menu level<br />

scroll down / scroll up<br />

1. Press the key<br />

The LCD DIALOG-DISPLAY shows the valid settings (see example above).<br />

2. Use the keys to select the menu item.<br />

3. Settings:<br />

Warn-Type: Select the parameter with the key<br />

enter<br />

(warning or alarm).<br />

and press<br />

Example:<br />

OverTemp and SubTemp: Use the numeral keypad to enter the value and<br />

press enter .<br />

SafeTemp: Set the new cut-out value using a screwdriver via the LCD<br />

DIALOG-DISPLAY. (Example: 50 °C)<br />

The value remains visible on the LCD DIALOG-DISPLAY for about 10<br />

seconds, or press<br />

to update the display immediately.<br />

27