Ultra-Low Refrigerated Circulators<br />

Return flow safety device<br />

ext SERIAL ALARM STAND-BY REG+<br />

Pt100<br />

E-PROG<br />

T16A<br />

T16A<br />

T1.25A<br />

13<br />

14<br />

Inputs/Outputs<br />

>SV-Out : Backflow<br />

Chan.1 : xxxxxxx<br />

Chan.2 : xxxxxxx<br />

(see page 50)<br />

If the liquid levels in the circulator bath and the external<br />

system are at different heights, overflowing must be<br />

prevented after the power has been turned off.<br />

Flood hazard!<br />

For this reason, solenoid valves for loop circuit or shut-off<br />

valves can be integrated in the loop circuit.<br />

Connect the valve to the connector (13).<br />

If socket (14) is used:<br />

In menu >Inputs/Outputs< set the<br />

menu item >SV-Out< to >BackflowControl< (see page 36).<br />

IMPORTANT: Measures for external control<br />

For external control and temperature measurement an external<br />

Pt100 sensor must be connected to the socket (17) on the rear of the<br />

circulator.<br />

Order No. Description Material cable<br />

8 981 003 200 x 6 mm dia., stainless steel 1.5 m<br />

8 981 005 200 x 6 mm dia., glass 1.5 m<br />

8 981 006 20 x 2 mm dia., stainless steel 1.5 m<br />

8 981 010 300 x 6 mm dia., stainless steel 1.5 m<br />

8 981 013 600 x 6 mm dia., stainless steel/Teflon 3 m<br />

8 981 014 1200 x 6 mm dia., stainless steel/Teflon 3 m<br />

8981103 Extension cable for Pt100 sensor 3.5 m<br />

8 981 020 M+R in-line Pt100 sensor<br />

The M+R in-line Pt100 sensor is a flow sensor and can be installed in<br />

the loop circuit.<br />

Pt100<br />

M+R<br />