Preparations<br />
3.4. Temperature application to external systems<br />
Temperature application to external, closed systems<br />
ext<br />
Pt100<br />
SERIAL ALARM STAND-BY REG+<br />
E-PROG<br />
The circulator is used for temperature application to external,<br />
closed systems (loop circuit) with simultaneous temperature<br />
application in the circulator bath.<br />
T16A<br />
T16A<br />
21 22<br />
T1.25A<br />
<br />
<br />
<br />
Unscrew the M16x1 collar nuts on the pump connectors with a<br />
19 mm (3/4“) wrench and remove the sealing disks. Using the<br />
collar nuts, screw on the tubing connection fittings (for tubing<br />
8 mm or 12 mm in diameter) delivered with the unit and tighten<br />
firmly. (Pressure pump: 22, suction pump: 21)<br />
Push on the tubings, and secure with tube clamps.<br />
Attach the tubing to the connectors of the external closed<br />
system, e.g., an instrument with a pressure-resistant<br />
temperature jacket or a temperature coil, and fasten with tube<br />
clamps to prevent slipping.<br />
Tubing see page 19<br />
Temperature application to external, open systems<br />
The circulator is equipped with both a pressure and suction pump<br />
for external temperature application in open systems.<br />
Differing flow rates of the pressure and suction pumps should be<br />
compensated. To maintain a constant liquid level, the JULABO<br />
„D+S“ Level Adapter is recommended for the external bath tank.<br />
The flow rate of the pressure pump will be then regulated by a<br />
built-in float device. The liquid level may be changed by a height<br />
adjustment on the „D+S“ Level Adapter.<br />
Accessory: „D+S“ Level Adapter Order No. 8 970 410<br />
S = Suction pump connection<br />
D = Pressure pump connection<br />
K = Float<br />
H = Height adjustment<br />
Important:<br />
The liquid level should be equal in the internal and external<br />
baths (absolute height).<br />
If you take out samples (for example Erlenmeyer flasks) from<br />
the external bath, turn the circulator off with the Start/Stop key.<br />
When working at temperatures below 0 °C and using the "D+S"<br />
Level Adapter do not use oil as the bath liquid. Condensing air<br />
will result in the formation of ice and thus affects the function of<br />
the float.<br />
Caution:<br />
Securely attach all tubing to prevent slipping.<br />
If the circulator is operated without external system, close the pump connector (22)<br />
with the cap nut.<br />