<strong>Operating</strong> controls and functional elements<br />

6 Adjustable excess temperature protection according to IEC 61010-2-010<br />

7<br />

8<br />

ext Pt100<br />

SERIAL<br />

Socket for external measurement and control sensor<br />

or external setpoint programming<br />

Interface RS232 / RS485: remote control via personal computer<br />

9 Socket: control cable of JULABO refrigerated circulator<br />

or output for alarm messages<br />

Option: Electronic module Order No. 8 900 100<br />

The circulator automatically recognizes the connected electronic module.<br />

10<br />

Alarm output (for external alarm signal)<br />

11<br />

12<br />

ALARM<br />

STAND-BY<br />

Standby input (for external emergency switch-off)<br />

Programmer input and temperature recorder output<br />

REG+E-PROG<br />

13 Connector for supplementary pump or solenoid valve<br />

230 V / max. 1.25 A<br />

No control voltage in the -OFF- condition<br />

14 Connector solenoid valve<br />

1. No control voltage in the -OFF- condition<br />

2. Configurable for controlled counter cooling via solenoid valve<br />

(only used with Heating Circulators)<br />

15 Fuses (for connectors 13, 14), T1.25A<br />

16a<br />

16b<br />

17a<br />

17b<br />

Mains fuses for circulator, T16A<br />

Mains fuses for cooling machine, T10A<br />

Mains power cable with plug for circulator<br />

Mains power cable with plug cooling machine<br />

18 F95, FW95 Built-in mains outlet for connection of circulator (230 V / 16 A)<br />

Built-in mains outlet 230 V / 6 A<br />

18 F81, F83, FP88 Built-in mains outlet for connection of circulator (230 V / 10 A)<br />

19 Drain port with drain tap<br />

20 Venting grid, removable<br />

21<br />

22 Pump connectors suction pump pressure pump<br />

OUT IN<br />

24<br />

Only for water cooled models: Cooling water OUTLET and INLET<br />