<strong>Operating</strong> manual<br />

Technical specifications<br />

F81-HL<br />

F83-HL<br />

Working temperature range °C -81 ... 100 -83 ... 50<br />

Temperature stability °C ±0,02 ±0,02<br />

Temperature selection<br />

digital<br />

via keypad<br />

indication on LCD DIALOG-DISPLAY (°C/°F)<br />

remote control via personal computer indication on monitor<br />

Temperature indication<br />

VFD COMFORT-DISPLAY (°C/°F)<br />

Resolution °C 0.01<br />

Absolute Temperature Calibration °C ±3<br />

Temperature control<br />

ICC - Intelligent Cascade Control<br />

Heater wattage (at 230 V) kW 1,3 1,3<br />

Cooling capacity °C +20 0 -20 -40 -60 -80 +20 0 -20 -40 -60 -80<br />

Medium ethanol kW 0.45 0.38 0.36 0.32 0.27 0.07 0.9 0.78 0.72 0.6 0.38 0.12<br />

Cooling compressor 2-stage 2-stage<br />

Refrigerant R404A/R23 R404A/R23<br />

Electronically adj. pump capacity stages 1 ... 4 1 ... 4<br />

Flow rate l/min at 0 bar 22 ... 26 22 ... 26<br />

Pressure max. bar at 0 liter 0.7 0.7<br />

Suction max bar at 0 liter 0.4 0.4<br />

Electrical connections:<br />

External alarm device 24-0 V DC / max. 25 mA<br />

Computer interface RS232 or RS485<br />

External Pt100 sensor<br />

Optional for HL and SL<br />

(Order No. 8900100 Electronic module with analog connections)<br />

Programmer input -100 °C to 400 °C = 0 - 10 V or 0 - 20 mA or 4 - 20 mA<br />

Input for the signal of a flow meter or external manipulated variable<br />

Temperature recorder outputs 0 - 10 V (0 V = -100 °C, 10 V = 400 °C)<br />

0 - 20 mA (0 mA = -100 °C, 20 mA = 400 °C)<br />

4 - 20 mA (4 mA = -100 °C, 20 mA = 400 °C)<br />

Standby input<br />

for external emergency switch-off<br />

Alarm output<br />

for external alarm signal<br />

Overall dimensions (WxDxH) cm 50x58x89 55x60x92<br />

Bath opening (WxL) cm 13x15 13x15<br />

Bath depth cm 16 16<br />

Filling volume liters 5 ... 6,5 5,5 ... 8<br />

Weight kg 88 118<br />

Ambient temperature °C 5 ... 40 5 ... 40<br />

Mains power connection V/ Hz 207-253 / 50-60 207-253 / 50<br />

230 V<br />

Current input (at 230 V) A 16 16<br />

Mains power connection V/ Hz<br />

207-253 / 60<br />

230 V/60 Hz<br />

Current input (at 230 V) A 16<br />

All measurements have been carried out at: rated voltage and frequency ambient temperature: 20 °C<br />