Appendix<br />

D<br />

Mobile Phase/Rinse Solvent<br />

• isocratic<br />

• 100% methanol<br />

Flow Rate<br />

0.7 mL/min<br />

<strong>Injection</strong><br />

Specifications Methods<br />

• 980 ng/mL anthracene in methanol<br />

• five times overfill total loop injection (100 µL)<br />

Rinse Volume<br />

• Probe at rinse station: aspirate 300 µL from<br />

reservoir at 15 mL/min then dispense 300 µL at<br />

16 mL/min<br />

• <strong>Injection</strong> port and valve: aspirate 450 µL from<br />

reservoir at 16 mL/min then dispense 450 µL at<br />

4 mL/min<br />

<strong>Injection</strong>s<br />

1 two system blank injections of methanol<br />

2 two initial blank injections of methanol<br />

3 one injection of 980 ng/mL anthracene<br />

4 four injections of methanol blanks<br />

Calibration Curve Parameters<br />

• external calibration<br />

• linear curve fit<br />

• five calibration levels<br />

1 0.049 ng/mL<br />

2 0.098 ng/mL<br />

3 0.294 ng/mL<br />

4 0.490 ng/mL<br />

5 0.980 ng/mL<br />

D-3