Appendix<br />
C<br />
Baud Rate<br />
To change the baud rate:<br />
1 Disconnect the calibrated tubing from the<br />
injection port. To fully remove the fitting, do the<br />
following:<br />
a) Loosen the two knurled screws that secure<br />
the injection module to the liquid handler.<br />
See diagram on page 2-2 if necessary.<br />
b) Lift and tilt the injection port plate from the<br />
top of the injection module with one hand<br />
while loosening the fitting with the other<br />
hand.<br />
2 Remove the four screws that secure the cover<br />
and then remove the cover.<br />
3 Locate the SW1 selector, on the left, near the<br />
back panel. This is the baud rate selector.<br />
4 Gently insert a small flat blade screwdriver into<br />
the numbered side of the SW1 selector and turn<br />
it.<br />
5 Align the white dot with one of the indicated<br />
numbers.<br />
6 Replace the cover and the four screws that<br />
secure it.<br />
7 Reconnect the calibrated tubing to the bottom<br />
of the injection port. You may need to lift and<br />
tilt the injection plate for better access to the<br />
injection port. Finger tighten the fitting. When<br />
secure, turn the fitting 1/4 turn with the 1/4"<br />
wrench.<br />
8 If necessary, re-install the injection port plate<br />
onto the injection module, making sure the<br />
plate is seated on the pins on the top of the<br />
module.<br />
9 Secure the injection module into place using the<br />
two knurled screws.<br />
C-2