Installation 2<br />

Adjusting the Liquid Handler’s Z-Arm Height<br />

If you are using a 215 SW Liquid Handler<br />

The 215 SW Setup Utility is supplied on the 215 SW<br />

Utility Programs CD-ROM supplied with your 215<br />

SW Liquid Handler.<br />

1 Start the 215 SW Setup Utility.<br />

2 To position the probe over the injection port,<br />

click Move to Port.<br />

3 To see if the probe’s Z height is aligned<br />

correctly, slide a small piece of paper between<br />

the probe’s tip and the injection port. The probe<br />

is aligned if the top of the paper catches the tip<br />

of the probe when you position the paper<br />

between the probe and the injection port.<br />

4 If the probe is not aligned, click on or to<br />

move the probe.<br />

5 When the probe is aligned, click Ok to store the<br />

new alignment in the liquid handler’s memory.<br />

5 Click on the About tab then Exit to close the<br />

215 SW Setup Utility.<br />

Note: If the probe requires adjustment along the X-<br />

or Y-axis, use the Adjust XY tab.<br />

2-18