Installation 2<br />

Adjusting the Liquid Handler’s Z-Arm Height<br />

Verification for <strong>Injection</strong> Port Use<br />

To ensure that the Z-arm is adjusted for use with<br />

the injection port, run INJECT_Z.EXE, the 215<br />

<strong>Injection</strong> Z-Height Utility, or the 215 SW Setup<br />

Utility.<br />

If you are using a Keypad Controller<br />

INJECT_Z.EXE is supplied on the 215 Keypad disk<br />

for use in a Gilson Keypad Controller.<br />

INJECT_Z.EXE prompts you:<br />

• to press a key so it can proceed with homing<br />

the liquid handler’s probe.<br />

• to specify the location of the <strong>819</strong> <strong>Injection</strong><br />

<strong>Module</strong>: next to the liquid handler’s right support<br />

or near the center of the liquid handler’s locator<br />

plate.<br />

• to adjust the height of the liquid handler’s<br />

probe so it is flush with the top of the injection<br />

port. To verify that the probe is on the same<br />

plane as the injection port, slide a small piece of<br />

paper (approximately 1" by 1") between the<br />

probe tip and injection port. The probe is aligned<br />

if the top of the paper catches the tip of the<br />

probe when it is slid between the probe and<br />

injection port.<br />

• to approve the adjustment made to the Z-arm<br />

height.<br />