Pipetman Concept Single-channel Models<br />

Currently not all models are available. Gilson Maximum ISO 8655 Maximum<br />

Consult your nearest local distributor for availability. Permissible Errors Permissible Errors<br />

Model Diamond Reference Volume Systematic Random Systematic Random<br />

w/o AC-adapter Tips * Number (μL) error (μL) error (μL) error (μL) error (μL)<br />

C10 • D10 F31012 0.5 ± 0.040 ≤ 0.013 ± 0.120 ≤ 0.080<br />

0.5 - 10 μL • DL10 1 ± 0.025 ≤ 0.012 ± 0.120 ≤ 0.080<br />

5 ± 0.060 ≤ 0.020 ± 0.120 ≤ 0.080<br />

10 ± 0.080 ≤ 0.025 ± 0.120 ≤ 0.080<br />

C100 • D200 F31013 5 ± 0.35 ≤ 0.10 ± 0.8 ≤ 0.30<br />

5 - 100 μL 10 ± 0.30 ≤ 0.10 ± 0.8 ≤ 0.30<br />

50 ± 0.38 ≤ 0.12 ± 0.8 ≤ 0.30<br />

100 ± 0.40 ≤ 0.15 ± 0.8 ≤ 0.30<br />

C300 • D300 F31014 20 ± 0.80 ≤ 0.16 ± 4.00 ≤ 1.50<br />

20 - 300 μL 30 ± 0.70 ≤ 0.20 ± 4.00 ≤ 1.50<br />

150 ± 0.90 ≤ 0.23 ± 4.00 ≤ 1.50<br />

300 ± 1.05 ≤ 0.30 ± 4.00 ≤ 1.50<br />

C1200 • D1200 F31015 100 ± 2.5 ≤ 0.4 ± 16.0 ≤ 6.0<br />

100 - 1200 μL 120 ± 2.4 ≤ 0.4 ± 16.0 ≤ 6.0<br />

600 ± 3.6 ≤ 0.8 ± 16.0 ≤ 6.0<br />

1200 ± 6.0 ≤ 1.2 ± 16.0 ≤ 6.0<br />

C5000 • D5000 F31016 500 ± 10 ≤ 2 ± 40 ≤ 15<br />

500 - 5000 μL 2500 ± 15 ≤ 4 ± 40 ≤ 15<br />

5000 ± 25 ≤ 7 ± 40 ≤ 15<br />

C10ml • D10ml F31017 1 mL ± 25 ≤ 4 ± 60 ≤ 30<br />

1 - 10 mL 5 mL ± 30 ≤ 8 ± 60 ≤ 30<br />

10 mL ± 50 ≤ 12 ± 60 ≤ 30<br />

* Recommended tip for establishing maximum permissible errors (ISO 8655 requirement). A large range of suitable Gilson Diamond tips is available. (See pages 22-27)<br />

Pipetman Concept Multichannel Models<br />

Currently not all models are available. Gilson Maximum ISO 8655 Maximum<br />

Consult your nearest local distributor for availability. Permissible Errors Permissible Errors<br />

Model Diamond Reference Volume Systematic Random Systematic Random<br />

w/o AC-adapter Tips * Number (μL) error (μL) error (μL) error (μL) error (μL)<br />

8x10 • DL10 F31032 1 ± 0.04 ≤ 0.02 ± 0.24 ≤ 0.16<br />

0.5 - 10 μL F31042 5 ± 0.08 ≤ 0.04 ± 0.24 ≤ 0.16<br />

12x10 10 ± 0.10 ≤ 0.06 ± 0.24 ≤ 0.16<br />

0.5 - 10 μL<br />

8x100 • D200 F31033 10 ± 0.25 ≤ 0.14 ± 1.60 ≤ 0.60<br />

5 - 100 μL F31043 50 ± 0.50 ≤ 0.20 ± 1.60 ≤ 0.60<br />

12x100 100 ± 0.80 ≤ 0.25 ± 1.60 ≤ 0.60<br />

5 - 100 μL<br />

8x300 • D300 F31034 30 ± 1.00 ≤ 0.18 ± 8.00 ≤ 3.00<br />

20 - 300μL F31044 150 ± 1.50 ≤ 0.38 ± 8.00 ≤ 3.00<br />

12x300 300 ± 2.40 ≤ 0.60 ± 8.00 ≤ 3.00<br />

20 - 300 μL<br />

* Recommended tip for establishing maximum permissible errors (ISO 8655 requirement). A large range of suitable Gilson Diamond tips is available. (See pages 22-27)