Manual Liquid Handling - John Morris Scientific
Manual Liquid Handling - John Morris Scientific
Manual Liquid Handling - John Morris Scientific
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
DISTRIMAN ® and DistriTips ®<br />
Simplify Repetitive Pipetting<br />
Distriman is a positive-displacement, continuously<br />
adjustable repeater-pipette to simplify multiple<br />
dispensing when used with Gilson DistriTip<br />
syringes.<br />
Direct volume<br />
readout<br />
With the Distriman you can dispense the exact volume,<br />
eliminating tedious calculations.<br />
Only 3 Syringes<br />
are needed<br />
Micro 125 µL, Mini 1250 µL and Maxi 12.5 mL cover the<br />
entire volume range (aliquots from 1 µL to 1.25 mL).<br />
Distriman is the only repetitive pipette available capable of<br />
dispensing more than 100 aliquots.<br />
Wide selection<br />
of packaging options<br />
Each model of DistriTip is available in standard or sterilized<br />
(ST) version, in boxes of 50 syringes. Sterilized DistriTips are<br />
individually wrapped and certified free of RNase, DNase,<br />
DNA, RNA and protease.<br />
Distriman<br />
Reference Number<br />
F164001<br />
DistriTips (box of 50 syringes)<br />
Model<br />
125 µL<br />
Micro<br />
Micro ST*<br />
1250 µL<br />
Mini<br />
Mini ST*<br />
12.5 mL<br />
Maxi<br />
Maxi ST*<br />
Aliquot<br />
Range<br />
1 µL to 12.5 µL<br />
10 µL to 125 µL<br />
100 µL to 1.25 mL<br />
Reference<br />
Number<br />
F164100<br />
F164130<br />
F164110<br />
F164140<br />
F164120<br />
F164150<br />
Volume<br />
(µL)<br />
Gilson Maximum<br />
Permissible Errors<br />
Systematic<br />
error (µL)<br />
Random<br />
error (µL)<br />
ISO 8655 Maximum<br />
Permissible Errors<br />
Systematic<br />
error (µL)<br />
Random<br />
error (µL)<br />
Min. 2 ± 0.100 ≤ 0.080 ±0.20 ≤ 0.10<br />
5 ±0.125 ≤ 0.075 ±0.20 ≤ 0.10<br />
Max. 10 ±0.200 ≤ 0.100 ±0.20 ≤ 0.10<br />
Min. 20 ±0.80 ≤ 0.20 ±1.50 ≤ 0.60<br />
50 ±1.00 ≤ 0.40 ±1.50 ≤ 0.60<br />
Max. 100 ±1.00 ≤ 0.60 ±1.50 ≤ 0.60<br />
Min. 200 ±6.00 ≤ 1.00 ±12.00 ≤ 4.00<br />
500 ±7.50 ≤ 1.50 ±12.00 ≤ 4.00<br />
Max. 1000 ±10.0 ≤ 2.50 ±12.00 ≤ 4.00<br />
*ST means sterilized syringes (individually wrapped).<br />
30<br />
<strong>Manual</strong> <strong>Liquid</strong> <strong>Handling</strong> Catalog 2007 / January - More information available at www.gilson.com