Test Phases<br />

Pretest Phase<br />

Overview<br />

The pretest phase consists of stages configured in each<br />

protocol. Commonly used stages are:<br />

n Supine<br />

n Standing<br />

Hyperventilating<br />

n<br />

Exercise Stress Test (Option): Test Phases<br />

Operating Steps<br />

1. Press the 12 ld to acquire and print a baseline <strong>ECG</strong>.<br />

1<br />

You can configure the Protocol Editor to take bloodpressure<br />

measurements manually or automatically. (See<br />

Chapter 7, “Edit Protocols Options”).<br />

The system will beep and display a message prompt in the<br />

header of the display.<br />

2<br />

A set of medians is saved at the end of the Pretest phase as<br />

baseline medians.<br />

Pretest<br />

st<br />

Pretest<br />

221C<br />

2. Press Pretest to advance to the next PRE-TEST stage.<br />

3. If you are using a treadmill to conduct the exercise test,<br />

tell the patient to place his/her feet on the treadmill<br />

frame, not on the belt.<br />

<strong>MAC</strong> <strong>5000</strong> <strong>System</strong> — 2000657-033, Revision A 6-5