Preparing the Patient: Apply the Electrodes<br />

Exercise Electrodes (with Exercise–Option)<br />

In addition to the standard electrodes, apply one electrode on the sternum (A1) and one in location V5R/C5R (A2).<br />

CM5, CC5, ML Lead Placement<br />

Electrode<br />

A1<br />

A2<br />

Electrode Placement<br />

Mid-sternum at the second intercostal space.<br />

In the fifth intercostal space in the anterior axillary line (V5R/C5R).<br />

254A<br />

<strong>MAC</strong> <strong>5000</strong> <strong>System</strong> — 2000657-033, Revision A 3-9