Preparing the Patient: Apply the Electrodes<br />

á To acquire a NEHB <strong>ECG</strong>, use the Standard 12 Lead electrode placement and items A and B shown below.<br />

Frank X,Y,Z Placement<br />

AHA Label IEC Label Electrode Placement<br />

A LA black L yellow Just below the clavicle of the right and left arms.<br />

B E orange E light blue Mid-sternum on the same horizontal level as C and D.<br />

H<br />

A<br />

C V4 blue C4 brown Mid-clavicular line in the fifth intercostal space.<br />

D V6 purple C6 purple Mid-axillary line on the same horizontal level as C.<br />

E LL red F green Right and left thighs.<br />

G<br />

B<br />

C D<br />

F RL green N black Right and left thighs.<br />

F<br />

E<br />

02B<br />

G I orange I light blue Right mid-axillary line on the same horizontal level as C and<br />

D.<br />

H RA white R red Just below the clavicle of the right and left arms.<br />

I H orange H light blue Back of neck, avoid the carotid artery and jugular vein.<br />

I<br />

J M orange M light blue Center of spine on the same horizontal level as C and D.<br />

J<br />

03B<br />

3-6 <strong>MAC</strong> <strong>5000</strong> <strong>System</strong> — 2000657-033, Revision A