Preparing the Patient: Apply the Electrodes<br />

Standard 15 Lead Placement<br />

AHA Label IEC Label Electrode Placement<br />

A V1 red C1 red Fourth intercostal space at the right sternal border.<br />

B V2 yellow C2 yellow Fourth intercostal space at the left sternal border.<br />

C V3 green C3 green Midway between location B and D.<br />

J<br />

K<br />

A B CD E F<br />

G<br />

D V4 blue C4 brown Mid-clavicular line in the fifth intercostal space.<br />

E V5 orange C5 black Anterior axillary line on the same horizontal level as D.<br />

F V6 purple C6 purple Mid-axillary line on the same horizontal level as D and E.<br />

I<br />

H<br />

088A<br />

G LA black L yellow Right and left deltoids.<br />

H LL red F green Right and left thighs.<br />

I RL green N black Right and left thighs.<br />

J RA white R red Right and left deltoids.<br />

K V4R gray C4R gray Right anterior chest opposite of D.<br />

L V8 gray C8 gray Under midscapular line.<br />

L M<br />

M V9 gray C9 gray Left paraspinal border.<br />

089A<br />

<strong>MAC</strong> <strong>5000</strong> <strong>System</strong> — 2000657-033, Revision A 3-5