Appendix D – Technical Specifications: Late Potential <strong>Analysis</strong><br />

Late Potential <strong>Analysis</strong><br />

(Hi-Res and PHi-Res Signal Averaged Electrocardiography)<br />

Item<br />

Frequency Response/Input<br />

Frequency Response/<br />

Output<br />

Upper Limit<br />

Lower Limit<br />

Sensitivities<br />

Raw Data and Template<br />

Average Beat<br />

Filtered Signals and<br />

Vector Magnitude<br />

<strong>Analysis</strong> Sampling Rate<br />

Digital Sampling Rate<br />

High/Low Pass Filters<br />

ADC Resolution<br />

<strong>Analysis</strong> Resolution<br />

-3 dB @ 0.01 and 250 Hz<br />

250 Hz<br />

0.01, 25, 40 or 80 Hz<br />

20 mm/mV<br />

20 mm/mV and 50 mm/mV<br />

1 mm/µV<br />

1000 samples per second per channel<br />

4000 samples per second per channel<br />

Spectral filter using Fast Fourier Transform (FFT)<br />

1.22 µV/LSB<br />

0.1525 µ V/LSB<br />

Description<br />