Computerized Electrocardiograph<br />

Appendix D – Technical Specifications: Computerized Electrocardiograph<br />

Item<br />

Instrument Type<br />

Battery Capacity<br />

<strong>ECG</strong> Storage<br />

Acquisition<br />

<strong>Analysis</strong><br />

<strong>Analysis</strong> Sampling Rate<br />

Digital Sampling Rate<br />

Dynamic Range<br />

Resolution<br />

Frequency Response<br />

Common-Mode Rejection<br />

Input Impedance<br />

Patient Leakage Current<br />

Pacemaker Detection<br />

Sensitivity<br />

Special Functions<br />

Description<br />

14 leadwire microcomputer-augmented, automatic electrocardiograph<br />

100 <strong>ECG</strong> acquisitions and one-page reports or 6 hours continuous operation (without printing)<br />

150 <strong>ECG</strong>’s (typ), 200 (max) on removable media (1.44 MB, 3.5” diskette)<br />

Simultaneous acquisition of up to 14 leadwires with programmable lead configurations<br />

12SL adult and 12SL pediatric analysis. Optional: HI-RES or PHI-RES late Potential <strong>Analysis</strong> (standard 12 leads plus leads<br />

V3R, V4R, V7), ACI-TIPI<br />

500 samples/s (sps)<br />

4000 samples/s/channel<br />

±320 mV DC, ±10 mV AC<br />

4.88 µV/LSB @ 500 sps<br />

–3 dB @ 0.01 to 150 Hz<br />

140 dB minimum (123 dB with AC filter disabled)<br />

>10 MW, defibrillator protected<br />