Appendix B – Troubleshooting: Equipment Problems<br />

<strong>System</strong> Errors<br />

The following errors may occur while you are operating this<br />

system. You may be required to perform some action.<br />

If you perform the recommended actions and the condition<br />

still remains, contact authorized service personnel. See<br />

“How to Reach Us” to find out how to contact GE <strong>Medical</strong><br />

<strong>System</strong>s Information Technologies.<br />

153A, 154A, 209A<br />

Problem Cause Solution<br />

appears on the screen. No battery is installed in the system. Install a battery and connect the system to an AC<br />

wall outlet to charge the battery.<br />

flashes intermittently. The battery charge is low. Connect the system to an AC wall outlet to charge<br />

the battery.<br />

appears on the screen. The writer door is open. Close the writer door.<br />

The system does not power up when operating<br />

from battery power.<br />

The system shuts down when operating from<br />

battery power.<br />

The battery is empty.<br />

Battery is empty, or the Automatic Shutdown<br />

feature is enabled.<br />

Connect the system to an AC wall outlet to charge<br />

the battery.<br />

Connect the system to an AC wall outlet to charge<br />

the battery, or power on the system.<br />

“X” Lead disconnected message appears. Electrode(s) disconnected. Reconnect the electrode(s).<br />

B-6 <strong>MAC</strong> <strong>5000</strong> <strong>System</strong> — 2000657-033, Revision A