Appendix B – Troubleshooting: Equipment Problems<br />

Equipment Problems<br />

Reducing <strong>ECG</strong> Data Noise<br />

If the acquired <strong>ECG</strong> data displays unacceptable noise<br />

levels:<br />

n Verify proper electrode placement.<br />

n Verify proper electrode application. Perspiration and<br />

dead skin cells must be removed from the electrode<br />

site.<br />

n Check for defective or date expired electrodes.<br />

n Check for defective, broken, or disconnected leadwires.<br />

n Check the patient’s position. The patient should remain<br />

motionless during the acquisition of a resting <strong>ECG</strong>.<br />

There is No ACI-TIPI Report<br />

n ACI-TIPI is disabled.<br />

u Enable ACI-TIPI.<br />

n<br />

u<br />

n<br />

u<br />

n<br />

n<br />

The selected report is without interpretation.<br />

Select Interpretation for the report.<br />

The ACI-TIPI required information is not entered.<br />

Make sure the age range, gender, and chest pain<br />

complaints are entered.<br />

The patient was entered as pediatric.<br />

u Make sure you enter an age range less than 16.<br />

The original <strong>ECG</strong> was acquired in an<br />

electrocardiograph without the ACI-TIPI option.<br />

B-4 <strong>MAC</strong> <strong>5000</strong> <strong>System</strong> — 2000657-033, Revision A