Appendix A – Maintenance: Maintaining the Battery<br />

Periodic Maintenance<br />

In addition to normal system use, periodic deep discharge<br />

cycles may be required to ensure consistent battery<br />

performance.<br />

A deep discharge cycle occurs when the battery is<br />

discharged until the system shuts down and the battery is<br />

charged until it is full.<br />

GE <strong>Medical</strong> <strong>System</strong>s Information Technologies<br />

recommends one deep discharge cycle once every three<br />

months.<br />

á GE <strong>Medical</strong> <strong>System</strong>s Information Technologies does not<br />

recommend over-exercising the battery with deep discharge<br />

cycles. See the <strong>MAC</strong> <strong>5000</strong> Service Manual for more battery<br />

maintenance and diagnostic information.<br />

Replacing the Battery<br />

:$51,1*<br />

Do NOT dispose of the battery pack by fire or<br />

burning. Follow local environmental guidelines<br />

concerning disposal and recycling.<br />

á If battery fluid contacts your skin, eyes, or clothing,<br />

immediately wash the area with clean water and see a doctor.<br />

á To prevent loss of data, remove the AC power cable from the<br />

AC wall outlet and power off the system before removing the<br />

battery.<br />

á After removing and replacing the battery, the battery gauge<br />

symbol resets to empty. Connect your system to an AC wall<br />

outlet to ensure a fully charged battery.<br />