Appendix A – Maintenance: Inspecting and Cleaning<br />

Cleaning<br />

Exterior Surfaces<br />

Clean the exterior surfaces of all equipment and peripheral<br />

devices monthly, or more frequently if needed.<br />

1. Use a clean, soft cloth and a mild dishwashing<br />

detergent diluted in water.<br />

2. Wring the excess water from the cloth. Do NOT drip<br />

water or any liquid on the writer assembly, and avoid<br />

contact with open vents, plugs, or connectors.<br />

3. Dry the surfaces with a clean cloth or paper towel.<br />

Disk Drive<br />

Clean the disk drive every six months, or more frequently if<br />

you have data storage problems.<br />

Use the head cleaning kit: PN 407546-001.<br />

á Do NOT use cleaning solutions or solvents on the disks. Data<br />

loss or disk drive failure could result.<br />

1. Power on the system.<br />

2. Follow the head cleaning kit instructions to prepare the<br />

head cleaning disk.<br />

3. Insert the head cleaning disk into the drive slot and<br />

note how long the drive spins.<br />

u The head cleaner disk must spin for a total of 30<br />

seconds to clean the drive.<br />

4. Repeat steps 1-3 until the head cleaning disk spins for<br />

30 seconds or more.<br />

5. Eject the head cleaning disk from the drive slot.<br />

<strong>MAC</strong> <strong>5000</strong> <strong>System</strong> — 2000657-033, Revision A A-5